SECTION 1 - SAFETY
2023491-201 Rev B
© 2007 General Electric Company
Page 7 of 33
CAUTION:
Possible Interference with Implanted Pacemaker
Therapy should not be delayed for patients with implanted pacemakers and a defibrillation attempt should be made
if the patient is unconscious and not breathing. The AED has pacemaker detection and rejection; however, with
some pacemakers the AED may not advise a defibrillation shock
1
.
Placing Pads:
• Do not place the pads directly over an implanted device.
• Place the pad at least one inch from any implanted device.
CAUTION:
Moving the Patient During a Rescue
During a rescue attempt, excessive jostling or moving of the patient may cause AEDs to improperly analyze the
patient’s cardiac rhythm. Stop all motion or vibration before attempting a rescue.
CAUTION:
Systems Statement
Equipment connected to the analog and digital interfaces must be certified to the respective IEC Standards (i.e.
IEC 60950 for data processing equipment and IEC 60601-1 for medical equipment). Furthermore, all
configurations shall comply with the system standard IEC 60601-1-1. Anybody who connects additional equipment
to the signal input part or signal output part configures a medical system, and is therefore, responsible that the
system complies with the requirements of the system standard IEC 60601-1-1.
CAUTION:
Case Cleaning Solutions
When disinfecting the case, use a non-oxidizing disinfectant, such as ammonium salts or a glutaraldehyde based
cleaning solution, to avoid damage to the metal connectors.
CAUTION:
The AED is programmed with software that has been tested to work with versions of ServiceLink and
RescueLink that are included with the AED. When using older version of ServiceLink and RescueLink are used to
communicate with this AED, there may be features described in this manual that are not available to be used.
Also, when communicating with an older AED with the version of ServiceLink and RescueLink included with this
new AED there may be features described in this manual that cannot be edited. The software in most cases will
give an error message when incompatibilities occur.