SECTION 4 – TECHNICAL DATA
2023491-201 Rev B
© 2007 General Electric Company
Page 33 of 33
RHYTHMX ECG ANALYSIS PERFORMANCE
The AED RHYTHMx ECG Analysis system analyzes the patient’s ECG and advises you when the AED detects a
shockable or non-shockable rhythm.
This system makes it possible for a person, with no training in the interpretation of ECG rhythms, to offer defibrillation
therapy to victims of sudden cardiac arrest.
CARDIAC RHYTHMS USED TO TEST THE RHYTHM RECOGNITION DETECTION SYSTEM
FOR CARDIAC SCIENCE AED
Rhythm Class
Specifications
Shockable Rhythm – VF
Meets AAMI DF 39 requirement and AHA recommendation of Sensitivity of >90%
Shockable Rhythm – VT
Meets AAMI DF 39 requirement and AHA recommendation of Sensitivity
of >75%
Non-Shockable Rhythm – NSR
Meets AAMI DF 39 requirement (>95%) and AHA recommendation (>99%) of
Specificity
Non-Shockable – Asystole
Meets AAMI DF 39 requirement and AHA recommendation of Specificity of >95%
Non-Shockable– all other rhythms
Meets AAMI DF 39 requirement and AHA recommendation of Specificity of >95%
Содержание 2023681
Страница 1: ...Service Manual 2023491 201rB ...