SECTION 2 - INTRODUCTION
2023491-201 Rev B
© 2007 General Electric Company
Page 15 of 33
BATTERY INSTALLATION
1. With the label on the battery facing the AED battery compartment, insert the
battery as shown in the drawing.
2. Push the latched end of the battery firmly into the AED, as shown in the
drawing, until the battery snaps into place. The exposed side of the battery
should be flush with the outside of the AED case.
3. Open the lid for 5 seconds to initiate self-test. If the battery is installed
properly, the STATUS INDICATOR will turn GREEN. Close the lid.
WARNING:
Battery Model 2023681-001 is Not Rechargeable
Do not attempt to recharge the battery. Any attempt to recharge the battery may result in an
explosion or fire hazard.
CAUTION:
Lithium Sulfur Dioxide Battery
Pressurized contents: Never recharge, short circuit, puncture, deform, or expose to temperatures
above 65°C (149°F). Remove the battery when discharged.
CAUTION:
Battery Disposal
Recycle or dispose of the lithium battery in accordance with all federal, state and local laws. To avoid
fire and explosion hazard, do not burn or incinerate the battery.
CAUTION:
Use only General Electric Approved Equipment
Using batteries, pads, cables, or optional equipment other than those approved by General Electric
may cause the Responder AED Pro to function improperly during a rescue.