Page 26 of 33
© 2007 General Electric Company
2023491-201 Rev B
WEIGHT
Model
Weight with Batteries and Pads
9300
3.20 kg (7.0 lb)
ENVIRONMENTAL OPERATION AND STANDBY CONDITIONS
Atmosphere Condition
Temperature
0°C to 50°C (32°F to 122°F)
Humidity
5% to 95% (non-condensing)
Pressure
57kPa (4,572m / +15,000ft) to 103kPa (-152m / -500ft)
SHIPMENT AND TRANSPORT ENVIRONMENTAL CONDITIONS (for up to 1 week)
Atmosphere Condition
Temperature
-30°C to 65°C (-22°F to 149°F)
Humidity
5% to 95% (non-condensing)
Pressure
57kPa (4,572m / +15,000ft) to 103kPa (-152m / -500ft)
PADS
•
Self-adhesive,
disposable defibrillation pads
•
Minimum combined surface area: 228cm2
•
Extended length of lead wire: 1.3m
LITHIUM BATTERY SPECIFICATIONS
•
Output voltage: 12VDC (max)
•
Batteries are non-rechargeable
•
Lithium contents: 9.2g (max)
•
Check local regulations for disposal information
Model
Estimated Shelf Life
Typical Shocks
2023681 Lithium
5 Years
Up to 290 shocks
The battery operating life depends on the type of battery, actual usage and environmental factors.
RECHARGEABLE BATTERY SPECIFICATIONS
•
Battery Voltage: 11.1V
•
Chemistry: Lithium-ion. Refer to local regulations.
•
Compatibility: Responder AED Pro Model 2023440
•
Battery Capacity: 60 shocks minimum (100 shocks typical) or 3 hours minimum (6 hours typical) of ECG display time.
•
Battery Charge Time: 3 hours for stated capacity, 4.5 hours to fully charge completely depleted battery.
•
Battery Standby: 6 months
•
Battery Life: 2.5 years or 300 Battery charge-discharge cycles, whichever comes first.
•
Battery Weight: 1 lb. 3 oz