SECTION 3 – MAINTENANCE AND TROUBLESHOOTING
2023491-201 Rev B
© 2007 General Electric Company
Page 23 of 33
ANNUAL MAINTENANCE
Perform the following tests annually to confirm that the diagnostics are functioning properly and to verify the integrity of
the case.
Check the Integrity of the Pads and Circuitry
1.
Open the AED lid.
2.
Remove the pads.
3.
Close the lid.
4.
Confirm that the
STATUS INDICATOR
turns
RED
.
5.
Open the lid and confirm that the
PAD
indicator is lit.
6.
Reconnect the pads and close the lid.
7.
Make sure the expiration date is visible through the clear window of the lid.
8.
Check to make sure that the
STATUS INDICATOR
is
GREEN
.
9.
Open the lid and confirm that no diagnostic indicators are lit.
10. Check the expiration date of the pads; if expired, replace them.
11. Check the pad’s packaging integrity.
12. Close the lid.
Check the Integrity of the Service Indicator (LED) and Circuitry
1.
Immediately after opening the AED lid, press and hold the
SHOCK
button and confirm that the
SERVICE
LED is lit.
2. Release
the
SHOCK
button.
3.
Close the lid.
4.
Verify that the
STATUS INDICATOR
remains
RED
.
5.
Open the lid and confirm that no diagnostic indicators are lit.
6.
Close the lid.
7. Verify
the
STATUS INDICATOR
turns
GREEN
.
Check the Integrity of the Case
Examine the molded case of the AED for any visible signs of stress. If the case shows signs of stress, contact Customer
Service or your local distributor.
Cleaning the AED Case
Gently clean the surface of the AED case with a damp sponge or with a cloth and mild soap.
CAUTION:
Case Cleaning Solutions
When disinfecting the case, use a non-oxidizing disinfectant, such as ammonium salts or a
glutaraldehyde based cleaning solution, to avoid damage to the metal connectors.
No periodic safety analysis tests referred to by the IEC 60601-1 international standard are
required.