Introduction
Before installing or using the device or system, be aware of the proximity of known RF
sources, such as the following:
• Radio and TV stations
• Portable and mobile RF communication devices (cell phones, two-way radios)
• X-ray, CT, or MRI devices
These devices are also possible sources of interference as they may emit higher
levels of electromagnetic radiation.
WARNING:
EQUIPMENT MALFUNCTION/INTERFERENCE — Use of portable phones or other
radio frequency (RF) emitting equipment near the system may cause unexpected
or adverse operation.
Do not use portable phones or other electronic equipment that may emit radio
frequency (RF) near this system.
WARNING:
EQUIPMENT MALFUNCTION/INTERFERENCE — Do not use the equipment or system
adjacent to, or stacked with, other equipment.
If adjacent or stacked use is necessary, test the equipment or system to verify
normal operation in the configuration in which you are using it.
WARNING:
ACCESSORIES/COMPONENTS — Adding accessories or components, or modifying
the medical device or system, may result in increased EMISSIONS or decreased
IMMUNITY of the device or system.
Use the following resources for more information on EMI/EMC and RF concerns:
• The Supplies and Accessories Reference Guide for your system
• Qualified GE Healthcare or approved third-party personnel
• The Electromagnetic Compatibility appendix in your system service or operator’s
manual
NOTE:
Compliance provides reasonable protection against radio-frequency interference.
However, there is no guarantee that interference will not occur in a particular
installation. You can tell whether this device or system is causing interference by
turning it off. If the interference stops, it was most likely caused by the device or
system.
Biocompatibility
The parts of the system described in this manual that come into contact with the
patient during the intended use, including all accessories, fulfill the biocompatibility
requirements of the applicable standards. If you have questions in this matter, contact
your GE Healthcare representative.
Legal Notice
GE Healthcare software contains several fields that can be filled in before performing
an ECG. Some of these fields are required, while others are optional and left to the
user to assess whether they are needed to perform the exam. The field
Race
is one
2053535-002K
MAC™ 2000
15
Содержание MAC 2000
Страница 30: ...Introduction 30 MAC 2000 2053535 002K ...
Страница 44: ...Product Overview 44 MAC 2000 2053535 002K ...
Страница 56: ...Preparing the Patient 56 MAC 2000 2053535 002K ...
Страница 62: ...Entering Patient Information 62 MAC 2000 2053535 002K ...
Страница 88: ...Arrhythmia Mode Recording 88 MAC 2000 2053535 002K ...
Страница 90: ...RR Analysis 90 MAC 2000 2053535 002K ...
Страница 94: ...RR Analysis 94 MAC 2000 2053535 002K ...
Страница 104: ...Stress Testing 104 MAC 2000 2053535 002K ...
Страница 182: ...Maintenance 182 MAC 2000 2053535 002K ...
Страница 212: ...Technical Specifications 212 MAC 2000 2053535 002K ...
Страница 224: ...224 MAC 2000 2053535 002K ...
Страница 225: ......