System Configuration
Resting ECG Setup Fields–Page 2 (cont'd.)
Field
Comment
ACI-TIPI
Enables/disables the inclusion of the ACI-TIPI
(Acute Cardiac Ischemia Time Insensitive Predictive
Instrument) statement and enables the
Chest Pain
field on the
Patient Information
window.
To include ACI-TIPI statements, the following
conditions must be met:
•
MI12
or
ME12
system option is activated
• TIPI system option is activated
• ACI-TIPI is enabled
• 10s ECG Report Format is enabled
• Print Interpretation is enabled
• Patient data includes: gender, date of birth, and
chest pain indication
• Patient cannot be a pediatric patient (15 years or
younger) as calculated form the date of birth
Sample Rate
Determines the report frequency. Options are
500
Hz
or
1000 Hz
. 1000 HZ is supported only for XML
output.
2053535-002K
MAC™ 2000
123