Technical Specifications
System Specifications
Instrument Type
Microprocessor augmented automatic electrocardiograph; 10-leadwire, 12 lead simultaneous acquisition
with programmable lead configuration.
Processing
Item
Specifications
ECG Interpretation
Marquette 12SL ECG Analysis Program for Adults and Pediatrics
Computerized Measurements
12-lead analysis
ECG Analysis Frequency
500 or 1000 samples/second/channel
Digital Sampling Rate
16000 samples/second/channel for normal data acquisition
Pace Sampling Rate
75K samples/second/channel
ECG On-screen Preview
On-screen preview of acquired 10 second ECG waveform
Acquisition Mode
Provides 10 seconds of instantaneous ECG acquisition
Dynamic Range
AC Differential ± 5 mV, DC offset ±300 mV
Resolution
4.88 µV +/-1% per LSB @ 500SPS
Frequency Range
0.04 to 150 Hz
Low Cutoff Frequency
0.04 Hz (ADS Off)
0.56 Hz (ADS On)
High Cutoff Frequency
Configurable at 20 Hz, 40 Hz, 100 Hz or 150 Hz
Common Mode Rejection
>135 dB (with 50/60 HZ filter ON)
Input Impedance
>10MΩ @ 10 Hz
Patient Leakage
<10 µA (Normal Condition), <50 µA (Single Fault Condition)
2053535-002K
MAC™ 2000
203
Содержание MAC 2000
Страница 30: ...Introduction 30 MAC 2000 2053535 002K ...
Страница 44: ...Product Overview 44 MAC 2000 2053535 002K ...
Страница 56: ...Preparing the Patient 56 MAC 2000 2053535 002K ...
Страница 62: ...Entering Patient Information 62 MAC 2000 2053535 002K ...
Страница 88: ...Arrhythmia Mode Recording 88 MAC 2000 2053535 002K ...
Страница 90: ...RR Analysis 90 MAC 2000 2053535 002K ...
Страница 94: ...RR Analysis 94 MAC 2000 2053535 002K ...
Страница 104: ...Stress Testing 104 MAC 2000 2053535 002K ...
Страница 182: ...Maintenance 182 MAC 2000 2053535 002K ...
Страница 212: ...Technical Specifications 212 MAC 2000 2053535 002K ...
Страница 224: ...224 MAC 2000 2053535 002K ...
Страница 225: ......