10
Instructions - 28-9414-37 Edition AA
Control Units CU-950 and CU-960
Figure 3-3.
Schematic UniNet-1 and mains cable connections for ÄKTApurifier.
How to connect the cables:
1
Connect the UniNet-1 data communication cable between the UniNet-1 port
on CU-950 and Pump P-900 in the system.
Note:
If a fraction collector is used, connect the fraction collector UniNet-1
cable between Pump P-900 and the fraction collector, and the 1.5 m
UniNet-1 cable between the fraction collector and CU-950, and then
the USB cable between CU-950 and the computer.
2
Connect the power converter cable to the Input port on CU-950.
3
Connect the mains cable between the power converter and the mains supply
socket at the rear of the system.
4
If an auto-sampler is used, the mains cable to the autosampler may be
connected to an external mains outlet if all mains supply sockets on the
system are occupied.
Note:
In this case the autosampler must be manually reset by disconnecting
the power cable when performing a system reset or restart.
5
Connect the USB cable to the USB port on CU-950. For connection to and
installation on the computer, please refer to the UNICORN Administration and
technical manual.
WARNING!
Only use mains cables delivered and approved by GE Healthcare.
Power
converter
Mains supply socket
Fraction
collector
(optional)
To computer
USB cable
UniNet-1 cable
UniNet-1 cable
CU-950
Mains cable