3
Installation
ÄKTAprime plus is delivered in protective packing material and shall be unpacked with
great care.
Any equipment connected to ÄKTAprime plus must fulfill applicable standards and local
regulations.
A video describing the installation process, is supplied with each ÄKTAprime plus system.
For detailed information on Installation, see
ÄKTAprime plus User Manual
.
3.1
Site requirements
Requirement
Parameter
Indoor use
Operation site
Maximum 2000 m
Altitude
Stable laboratory bench e.g. 120 × 80 cm
Placement
100-120/220-240 V AC ±10%, 50-60 Hz
Electrical power
Overvoltage category II
Transient overvoltages
4°C to 40°C
Ambient temperature
20% to 95%, non-condensing
Humidity
84 to 106 kPa (840 to 1060 mbar)
Atmospheric pressure
2
Pollution degree
WARNING
Do not block the ventilation inlets or outlets on the system.
3.2
Transport
The equipment can be transported on a trolley capable of supporting at least 20 kg.
ÄKTAprime plus Operating Instructions 28-9597-89 AC
29
3 Installation