WARNING
Before disassembly, check that there is no pressure in the piping
system.
WARNING
After assembly, the piping system must be tested for leakage at
maximum pressure for continued protection against injury risks
due to fluid jets, burst pipes or explosive atmosphere.
5.6
Calibration
The table below lists the type and frequency of calibrations that can be done on the in-
strument. Refer to PrimeView user documentation and to the individual component User
Manuals and Instructions for descriptions of how to perform these calibrations. The
calibrations are performed from PrimeView by selecting
System:Calibrate
in
System
Control
.
How often
Component
Every day.
pH monitor (if applicable)
When required.
Pump (if applicable)
When required.
Pressure reading
Only necessary if specific conductivity with high
accuracy is measured (
Cond_Calib
).
Cell constant
Conductivity
flow cell
Must be done when changing the conductivity
flow cell (
Temp
).
Temperature
Must be done when changing the conductivity
flow cell (
Cond_Cell
).
Entering a
new cell con-
stant