In  accordance  with  the  EMC-regulations  for  medical  products  we  are  obliged  by  law  to 
provide the following information. 

Guidance and manufacturer’s declaration — electromagnetic emissions 

The  equipment  is  intended  for use in the electromagnetic environment specified below. The customer or 
the user of the equipment should assure that it is used in said environment. 

Emissions test 

Compliance 

Electromagnetic environment – guidance 

RF emissions, 
CISPR 11 

Group 1 

The  equipment  uses  RF  energy  only  for  its  internal  function. 
Therefore,  its RF emissions are very low and are not likely to cause 
any interference in nearby electronic equipment. 

RF emissions, 
CISPR 11 

Class B 

The  equipment  is  suitable  for  use  in  all  establishments,  including 
domestic  establishments  and  those  directly  connected  to  the  public 
low-voltage  power  supply  network  that  supplies  buildings  used  for 
domestic purposes. 

Harmonic emissions, 
IEC 61000-3-2 (*) 

Class A 

Voltage fluctuation/flicker 
emissions, IEC 61000-3-3 (*) 

Complies 

(*) Note: For devices with a power consumption between 75 W and 1000 W only. 

Guidance and manufacturer’s declaration — electromagnetic immunity 

The  equipment  is  intended  for use in the electromagnetic environment specified below. The customer or 
the user of the equipment should assure that it is used in said environment. 

Immunity test 

IEC 60601- test level 

Compliance level 

Electromagnetic environment – 

guidance 

Electrostatic discharge (ESD), 
IEC61000-4-2 

6 kV  contact 

 

8 kV  air 

6 kV  contact 

 

8 kV  air 

Floors should be wood, concrete or 
ceramic tile. If floors are covered with 
synthetic material, the relative humidity 
should be at least 30 %. 

Electrical fast transient/burst,  
IEC 61000-4-4 

2 kV for power supply 

lines  
 

1 kV for input/output 

lines 

2 kV for power supply 

lines  
 

1 kV for input/output 

lines 

Mains power quality should be that of a 
typical commercial or hospital 
environment. 

Surge,  
IEC 61000-4-5 

1 kV differential mode 

 

2 kV common mode 

1 kV differential mode 

 

2 kV common mode 

Mains power quality should be that of a 
typical commercial or hospital 
environment. 

Voltage dips, short 
interruptions and voltage 
variations on power supply 
input lines,  
IEC 61000-4-11 
 

<5 % U  
for ½ cycle 
(>95 % dip) 
 
40 % U  
for 5 cycles 
60 % dip) 
 
70 % U  
for 25 cycles 
30 % dip) 
 
<95 % U  
for 5 s 
(>5 % dip) 
 

<5 % U  
for ½ cycle 
(>95 % dip) 
 
40 % U  
for 5 cycles 
60 % dip) 
 
70 % U  
for 25 cycles 
30 % dip) 
 
<95 % U  
for 5 s 
(>5 % dip) 
 

Mains power quality should be that of a 
typical commercial or hospital 
environment. 
 
If the user of the equipment requires 
continued operation during power mains 
interruptions, it is recommended that the 
equipment be powered from an 
uninterruptible power supply or a 
battery. 

Power frequency (50/60 Hz) 
magnetic field,  
IEC 61000-4-8 

3 A/m 

3 A/m 

Power frequency magnetic fields should 
be at levels characteristic of a typical 
location in a typical commercial or 
hospital environment. 

Note: U  is the a.c. mains voltage prior to application of the test level. 

 

Содержание TESI Relax

Страница 1: ...gbo Medizintechnik AG Version 2 1 Massage Table TESI Relax User Manual...

Страница 2: ...omissions All rights reserved No part of this manual may be reproduced in any form or by any means electronic mechanical or otherwise without the prior written permission of gbo Medizintechnik AG gbo...

Страница 3: ...eans Caution Observe the instructions for use Note Information that will facilitate your work Glossary Touch Screen Display equipped with a touch panel The touch panel reacts to the smooth touch of th...

Страница 4: ...nection and switch on 10 2 2 1 Fuses 11 2 3 Placing out of operation 11 2 4 Waste removal of the device and accessories 11 2 5 Settings 12 2 5 1 The menu 12 2 5 2 Device settings 12 2 5 3 Basic settin...

Страница 5: ...21 5 1 Notes 21 5 2 Warnings 21 5 3 Errors 22 5 4 Other notes 22 6 MAINTENANCE 23 6 1 Legal provisions and requirements 23 6 2 Safety controls 23 6 3 Cleaning disinfection and care 24 7 WARNINGS AND S...

Страница 6: ...l massage vibration The system permits the optional or combined application of these modes While the patient feels very comfortable with the treatment for the therapist it offers high operational comf...

Страница 7: ...ard 1 Slot for Chip Card 2 Leg rest 9 Optional connector for support arm left 10a Lock for service access 11 Head rest 13 Figure 1 Device view Middle section Massage section 12 Stepped bed 3 Support a...

Страница 8: ...ing the height press the OK Button and the treatment will be started with the basic settings If you use the remote control Press the button to activate NO Press OK The request to adjust the height of...

Страница 9: ...elax 9 gbo Medizintechnik AG Version 2 1 Note You can handle the table at any time with the Touch Screen as well as with the remote control The activated function will appear in green colour on the di...

Страница 10: ...eas of anesthesia departments the possibility of an explosion cannot be excluded If the patient and or the patient cable is directly exposed to a radiator of a medical device for high frequency heat t...

Страница 11: ...3 With a screw driver the box can be pulled out of the receiver by the small slot 4 Only fuses that correspond to the specifications named in the technical data should be used Warning Risk of fire if...

Страница 12: ...ed on and is ready you can start the threapy immediately using the customized settings Note You can only start the menu if no treatment is in progress and the correct chip card is inserted You can han...

Страница 13: ...stored when the device starts up Settings Possible options Basic settings upon delivery Compression of the rolls 0 100 50 Speed cm s 0 15 5 Vibration Hz off 15 35 19 Spine area analog Height of steppe...

Страница 14: ...time account You can choose between 1 2 5 and 10 hours Confirm your choice with the OK Button 7 Eject the service or adjustment card 8 Insert the card that you want to create in the intended slot at...

Страница 15: ...and the relating meaning Type of signal Reason Information sound In case of a failure Triad gong At the end of the faultless automatic check when switching on the device At the end of the treatment ti...

Страница 16: ...etting Pressure of the rolls 0 100 Speed cm s 0 15 cm s Vibration Hz off 15 35 Hz Spine area analog Height of stepped bed cm 0 50 cm Temperature C 20 35 C Time of treatment min 1 30 min Volume of audi...

Страница 17: ...r the next treatment 4 Card with time credit With this chip card the patient can use the TESI Relax for pre paid time of treatment However no treatment settings can be stored on this card 3 6 Head pho...

Страница 18: ...red Raise of step 2 Touch the Button for massage The roller pressure will be set on 50 3 Touch the Buttons to set the treatment time 4 Touch the Buttons to set the roller pressure 5 Touch the Buttons...

Страница 19: ...NO on the Touch Screen or by pressing the Button Button NO will be activated pressing the Button OK Continue with point 5 b b Alternatively If you use a new personal card or card with time credit you...

Страница 20: ...rd TESI Relax will be deactivated 10 After the treatment the patient should remain in an upright position with his legs dangling over the side of the table for about 3 to 5 minutes and remain quiet an...

Страница 21: ...finger at a single point If more than one finger or the ball of the thumb are placed on the active surface this will trigger the error message shown above 2 Confirm the message with the Button OK Not...

Страница 22: ...hing the Button OK and try to move the carriage again by touching the Touch Screen 3 If the message appears again contact the Service Center 5 3 Errors There are two possible error messages Error CAN...

Страница 23: ...user 6 1 Legal provisions and requirements The device is subject to the provisions of the Medical Device Directive The safety controls have to be carried out on the basis of this directive Thereby th...

Страница 24: ...portions of alcohol Example for usefull products to be used for cleaning are Mikrobac Tissues by Hartmann Bode or Cleanisept Wipes by Dr Schumacher If the imitation leather of the heating cover is we...

Страница 25: ...signs used 0123 CE Conformity sign Observe the instructions for use Application section with protection class B This product complies with WEEE Directive 2002 96 EG waste electrical and electronic eq...

Страница 26: ...f the support arm can be used to get up Dimensions max 66 175 cm x 200 cm x 94 cm height x depth x width height of the device without with support arm Weight max 80 kg without accessories Color RAL 90...

Страница 27: ...lster small Compact flash card Relaxation CD Head phones set 018 0 0010 Chip card freely programmable 018 4 0006 Adjustment card 018 4 0006 EK Infrared remote control 018 1 0059 Head phones 018 4 0012...

Страница 28: ...of treatment 19 Creating chip cards 14 D Device settings 12 F Fuses 11 I Information sound 15 M Maintenance 23 massage rollers 18 N Note 8 9 12 14 18 20 R Remote control 15 S settings 12 Short instruc...

Страница 29: ...nical support by the manufacturer dealers or service authorized by the manufacturer The product s life time anticipated by the manufacturer is 10 years The TESI Relax tables are electronic devices For...

Страница 30: ...est IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ce...

Страница 31: ...z to 800 MHz d 2 3 P for 800 MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts according to the transmitter manufacturer and d is the recommended separation distanc...

Страница 32: ...32 TESI Relax gbo Medizintechnik AG Version 2 1...

Отзывы: