background image

TESI

 

Relax 

19

 

 

 gbo Medizintechnik AG

 

Version 2.1

 

 

4.3

 

Course of treatment 

In general, the course of treatment is as follows: 
  

 

Figure 3: 

 Basic display of Touch Screens            

 

 

1.

 

Follow the instructions on the Touch Screen. 

 

2.

 

On the Touch Screen you find the request: „Please insert your chip card to activate!“ 

 

3.

 

Insert the chip card into the slot. The table will be unlocked. 

 

4.

 

The patient positions himself on the table according to the intended treatment. 

 
5.

 

a) 

In case of current settings stored,

 on the Touch Screen you will find the request: „Load settings of 

the last treatment from chip card?” 

 

By  touching  the 

Button  (YES)

  on  the  Touch  Screen  or  by  pressing  the 

Button

 

(OK)

  on  the 

remote control the treatment of the stored settings will be started. Continue with point 6. 

 

By touching the 

Button (NO)

 on the Touch Screen or by pressing the 

Button (+) 

→ 

Button (NO)

 

will be activated → pressing the 

Button (OK).

 Continue with point 5.b)  

b) 

Alternatively:

  If  you  use  a  new  personal  card  or  card  with  time  credit  you  will  find  the  request: 

“Please change height of stepped bed using the keys +‘ and ‚

-

‘ . Proceed with ‘OK’.“ 

Change  height  of  stepped  bed  accordingly  and  confirm  with 

OK. 

  Treatment  will  start  with  basic 

settings. 

 

6.

 

If the patient wants some changes in the settings, he can adjust them at any time during the treatment by 
the  remote  control.  Optionally,  the  settings  can  be  changed  on  the  Touch  Screen  by  touching  the 
respective buttons. 

 

7.

 

The treatment time ends with the triad-gong . The massage rollers will move down and stop. Vibration 
stops.

 

  

8.

 

If  a 

personal  chip  card

  is  used,  current  settings  of  the  treatment  will  be  stored  at  the  end  of  the 

treatment to make it possible to use them for the next treatment. 

 

User guidance

 and hints 

for 

remote control

 

Patient has to follow the 

instructions. 

Содержание TESI Relax

Страница 1: ...gbo Medizintechnik AG Version 2 1 Massage Table TESI Relax User Manual...

Страница 2: ...omissions All rights reserved No part of this manual may be reproduced in any form or by any means electronic mechanical or otherwise without the prior written permission of gbo Medizintechnik AG gbo...

Страница 3: ...eans Caution Observe the instructions for use Note Information that will facilitate your work Glossary Touch Screen Display equipped with a touch panel The touch panel reacts to the smooth touch of th...

Страница 4: ...nection and switch on 10 2 2 1 Fuses 11 2 3 Placing out of operation 11 2 4 Waste removal of the device and accessories 11 2 5 Settings 12 2 5 1 The menu 12 2 5 2 Device settings 12 2 5 3 Basic settin...

Страница 5: ...21 5 1 Notes 21 5 2 Warnings 21 5 3 Errors 22 5 4 Other notes 22 6 MAINTENANCE 23 6 1 Legal provisions and requirements 23 6 2 Safety controls 23 6 3 Cleaning disinfection and care 24 7 WARNINGS AND S...

Страница 6: ...l massage vibration The system permits the optional or combined application of these modes While the patient feels very comfortable with the treatment for the therapist it offers high operational comf...

Страница 7: ...ard 1 Slot for Chip Card 2 Leg rest 9 Optional connector for support arm left 10a Lock for service access 11 Head rest 13 Figure 1 Device view Middle section Massage section 12 Stepped bed 3 Support a...

Страница 8: ...ing the height press the OK Button and the treatment will be started with the basic settings If you use the remote control Press the button to activate NO Press OK The request to adjust the height of...

Страница 9: ...elax 9 gbo Medizintechnik AG Version 2 1 Note You can handle the table at any time with the Touch Screen as well as with the remote control The activated function will appear in green colour on the di...

Страница 10: ...eas of anesthesia departments the possibility of an explosion cannot be excluded If the patient and or the patient cable is directly exposed to a radiator of a medical device for high frequency heat t...

Страница 11: ...3 With a screw driver the box can be pulled out of the receiver by the small slot 4 Only fuses that correspond to the specifications named in the technical data should be used Warning Risk of fire if...

Страница 12: ...ed on and is ready you can start the threapy immediately using the customized settings Note You can only start the menu if no treatment is in progress and the correct chip card is inserted You can han...

Страница 13: ...stored when the device starts up Settings Possible options Basic settings upon delivery Compression of the rolls 0 100 50 Speed cm s 0 15 5 Vibration Hz off 15 35 19 Spine area analog Height of steppe...

Страница 14: ...time account You can choose between 1 2 5 and 10 hours Confirm your choice with the OK Button 7 Eject the service or adjustment card 8 Insert the card that you want to create in the intended slot at...

Страница 15: ...and the relating meaning Type of signal Reason Information sound In case of a failure Triad gong At the end of the faultless automatic check when switching on the device At the end of the treatment ti...

Страница 16: ...etting Pressure of the rolls 0 100 Speed cm s 0 15 cm s Vibration Hz off 15 35 Hz Spine area analog Height of stepped bed cm 0 50 cm Temperature C 20 35 C Time of treatment min 1 30 min Volume of audi...

Страница 17: ...r the next treatment 4 Card with time credit With this chip card the patient can use the TESI Relax for pre paid time of treatment However no treatment settings can be stored on this card 3 6 Head pho...

Страница 18: ...red Raise of step 2 Touch the Button for massage The roller pressure will be set on 50 3 Touch the Buttons to set the treatment time 4 Touch the Buttons to set the roller pressure 5 Touch the Buttons...

Страница 19: ...NO on the Touch Screen or by pressing the Button Button NO will be activated pressing the Button OK Continue with point 5 b b Alternatively If you use a new personal card or card with time credit you...

Страница 20: ...rd TESI Relax will be deactivated 10 After the treatment the patient should remain in an upright position with his legs dangling over the side of the table for about 3 to 5 minutes and remain quiet an...

Страница 21: ...finger at a single point If more than one finger or the ball of the thumb are placed on the active surface this will trigger the error message shown above 2 Confirm the message with the Button OK Not...

Страница 22: ...hing the Button OK and try to move the carriage again by touching the Touch Screen 3 If the message appears again contact the Service Center 5 3 Errors There are two possible error messages Error CAN...

Страница 23: ...user 6 1 Legal provisions and requirements The device is subject to the provisions of the Medical Device Directive The safety controls have to be carried out on the basis of this directive Thereby th...

Страница 24: ...portions of alcohol Example for usefull products to be used for cleaning are Mikrobac Tissues by Hartmann Bode or Cleanisept Wipes by Dr Schumacher If the imitation leather of the heating cover is we...

Страница 25: ...signs used 0123 CE Conformity sign Observe the instructions for use Application section with protection class B This product complies with WEEE Directive 2002 96 EG waste electrical and electronic eq...

Страница 26: ...f the support arm can be used to get up Dimensions max 66 175 cm x 200 cm x 94 cm height x depth x width height of the device without with support arm Weight max 80 kg without accessories Color RAL 90...

Страница 27: ...lster small Compact flash card Relaxation CD Head phones set 018 0 0010 Chip card freely programmable 018 4 0006 Adjustment card 018 4 0006 EK Infrared remote control 018 1 0059 Head phones 018 4 0012...

Страница 28: ...of treatment 19 Creating chip cards 14 D Device settings 12 F Fuses 11 I Information sound 15 M Maintenance 23 massage rollers 18 N Note 8 9 12 14 18 20 R Remote control 15 S settings 12 Short instruc...

Страница 29: ...nical support by the manufacturer dealers or service authorized by the manufacturer The product s life time anticipated by the manufacturer is 10 years The TESI Relax tables are electronic devices For...

Страница 30: ...est IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ce...

Страница 31: ...z to 800 MHz d 2 3 P for 800 MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts according to the transmitter manufacturer and d is the recommended separation distanc...

Страница 32: ...32 TESI Relax gbo Medizintechnik AG Version 2 1...

Отзывы: