51
1
Open dialog:
System
›
Settings
›
Current
Loop
>
Settings…
.
2
With
Result Value / Special
Signals -> Current
›
Min
(
1
)
and
Max
(
2
) set the mapping
of the concentration range
(
3
) to the range 0 mA
–
20
mA.
Here:
0.1 mg/m
3
→
4.0 mA
7.3 mg/m
3
→
20.0 mA
3
With
Result Value / Special
Signals -> Current
›
DAC [0
mA]
and
DAC [20 mA]
calibrate to 0-20 mA by
measuring the output Value
in mA on an external display
unit. The values in the
Screenshot
are
preset
factory default values and
have to be finetuned by the
user.