32
8
Device Parameters During
Measurements and Standby
8.1
Introduction
During each measurement and calibration device component values are
dynamically modified. These values are displayed in the scope view.
Adjusting values are displayed as red.
The values listed below may vary slightly from the actual values due to
customization.
The GC-IMS-Silox is designed for routine measurements to quantify
Siloxanes.In between measurements the Instrument is set to a Standby
mode for minimizing Nitrogen consumption
8.2
Values
Device
Component
Values
Standby
MEA
CAL
INTERVAL
T1 [IMS]
65 °C
65 °C
65 °C
65 °C
T2 [GC]
80 °C
80 °C
80 °C
80 °C
T3 [Sample Loop]
70 °C
70 °C
70 °C
70 °C
T5 [Transfer Line]
80 °C
80 °C
80 °C
80 °C
EPC1 [Drift Gas]
10 ml/min
150 ml/min
150 ml/min
10/150 ml/min
EPC2 [Carrier Gas]
10 ml/min
5-15 ml/min
5-15 ml/min
5-15 ml/min
Содержание GC-IMS-SILOX
Страница 1: ...G A S Gesellschaft f r analytische Sensorsysteme mbH GC IMS SILOX USER MANUAL...
Страница 66: ...65 7 Close the dialog with Close The import process is completed...
Страница 103: ...102 17 10 2 System Page Plan Tab Figure 22 The System Page Plan...
Страница 128: ...127 Exemplary Readout on Modbus...
Страница 136: ...135 Figure 31 Results Window...