38
10
Workflow: Manual
Calibration
10.1
Introduction
The purpose of the manual calibration is to fine-tune the calibration the device.
One or more substance calibrations are modified in the calibration
process. Make sure that an appropriate calibration gas source is
connected to the Testgas In socket at the rear of the housing.
A manual calibration only works within certain margins around the
factory calibration. When it varies too much from the factory
calibration the user is informed, and the results of the manual
calibration process are discarded.
10.2
Workflow
Workflow 4: Manual Calibration
During a measurement device component values are dynamically
modified. These values are displayed in the scope view. See
Parameters During Measurements and
Any cleaning process must be completed ~2 hours prior to any
measurement so that the internal components can cool down to the
required temperatures.