GASCOSEEKER 2-500 USER HANDBOOK
70
GAS RESPONSE IS SLOW
IS THE INSTRUMENT DETECTING METHANE?
The instrument response is setup using METHANE. If you
are detecting other substances such as BUTANE. PETROL
VAPOUR etc, the response may appear slow.
ACTION:
CHECK YOU HAVE CORRECT
INSTRUMENT FOR THE GAS YOU ARE
USING.
IF IN DOUBT ASK SUPERVISOR.
ARE THE BAT OR BAT FAULT FLAGS ACTIVATED?
If the batteries are low, the pump flow rate may be reduced
and hence the response time reduced.
ACTION:
CHANGE BATTERIES FOR NEW LR20
ALKALINE CELLS.
IS THE SAMPLE LINE OR PROBE TRAPPED OR
BLOCKED?
If the sample line or filters are blocked, the gas sample
rate may be reduced.
ACTION:
CHECK FILTERS AND SAMPLE LINE
FOR BLOCKAGE, RESTRICTION OR
CONTAMINATION.
REPLACE AS NECESSARY.