GASCOSEEKER 2-500 USER HANDBOOK
22
OPERATOR MESSAGES
A number of messages can appear in the display window,
indicating instrument status and operator action.
READ AND UNDERSTAND HANDBOOK BEFORE USE
SAMPLE
Indicates that the pump is running and the instrument is
sampling.
OFF
Indicates that the instrument is about to switch OFF
Can be reset by pressing MEASURE button once.
SAMPLE FAULT
Indicates that the instrument sample flow is incorrect due
to sample path blockage or pump failure.
CHECK ZERO
Indication that there may have been a zero shift due to the
presence of gas. Switch off the instrument and switch on
again in fresh air.
ZERO FAULT
Indicates that zero is outside calibration limits.
Switch instrument OFF and initiate ZERO sequence with
instrument in fresh air to reset zero.
BAT ‘Flashing’
Indicates that the batteries will soon require replacement
as indicated by BAT FAULT flag.
BAT FAULT
Indicates that the batteries require replacement.
ZERO
Indicates that the instrument is in the ZERO mode.