GASCOSEEKER 2-500 USER HANDBOOK
62
HAS INSTRUMENT BEEN USED TO MEASURE
PETROL, BUTANE OR SOME OTHER SUBSTANCE
OTHER THAN METHANE?
If instrument has been used for a heavy hydrocarbon or a
volatile substance such as Petrol, this may be impregnated
into the filters or some residue may be in sample line,
which will register as a false zero in the zero check
sequence.
ACTION:
CHECK FILTERS FOR CONTAMINATION.
CHANGE. IF NECESSARY.
REMOVE SAMPLE LINE AND PROBE.
SWITCH INSTRUMENT OFF.
INITIATE A ZERO SEQUENCE IN FRESH
AIR.
HAS INSTRUMENT BEEN SWITCHED OFF IN GASEOUS
ATMOSPHERE?
If the instrument was switched off with gas in the sample
path then some may remain and this would alter the zero
setting, which would be detected by the instrument and
indicated as a ZERO FAULT.
ACTION:
FLUSH INSTRUMENT WITH FRESH AIR
SWITCH OFF
INITIATE A ZERO SEQUENCE IN FRESH
AIR.