Gas Detection (Australia) Pty Ltd Page 22 of 44
M-GDA-3000 Rev 1
9.
Menu Configuration Table
LEVEL1
LEVEL2
DEFAULT
NAME
PARAMETER
PROGRAMMABLE
MODE
GROUP OF GAS SEL
(GROUP OF GAS
SELECT)
HC/PROPANE/CO/CO2
HC
UNIT & TAG SEL.
(UNIT & TAG SELECT)
%/%LEL/PPM/PPB
%LEL
DECIMAL POINT
0.100/1.00/10.0/100
100
HIGH SCALE ADJ.
(HIGH SCALE ADJUST)
1~10000
100
PASSWORD SET
00~99
00
CALIBRATION
MODE
CALIBRATION
[ZERO & SPAN]
ZERO CALIBRATION
[NO,YES]
[NO]
ZERO GAS
[ 0]
ZERO PROCESSING
[SUCCESS / FAIL]
CALIBRATION DATA
[ 0]
SPAN CALIBRATION
[NO , YES]
[NO]
SPAN GAS VALUE
[ 0]
50%
of Full Scale
[CH4] SPAN GAS
000 [ 0]
SPAN PROCESSING
[SUCCESS / FAIL ]
CALIBRATION DATA
[ 0]
ALARM PROGRAM
MODE
ALARM OPERATING
[AUTO/MANUAL]
AUTO
ALARM RELAY TYPE
DE-
ENERGIZED/ENERGIZE
D
DE-
ENERGIZED
FAULT RELAY TYPE
DE-
ENERGIZED/ENERGIZE
D
DE-
ENERGIZED
ALARM1 TYPE SEL.
(ALARM1 TYPE SELECT)
[INCREASE/
DECREASE]
INCREASE
ALARM1 LEVEL ADJ
(ALARM1 LEVEL ADJUST)
1~Full Scale ADJ
20
ALARM1 DEAD BAND
0.0~ Adj within 10%
of Full Scale
1.0