Multi Gas Clip Simple Plus User’s Manual

UM MGC-S-PLUS v1.00a 

Portable gas detectors you can count on

1 of 18

WARNING STATEMENTS/AVERTISSEEMENT

 Do not substitute components as this may interfere with the intrinsic safety of the device .
 Ne remplacez pas les composants car cela pourrait interférer avec la sécurité intrinsèque de l’appareil .
 DO NOT substitute any other battery type than specified and supplied by Gas Clip Technologies . 
 Only use Gas Clip Technologies parts in the detector . Non-approved parts will void the warranty and are 

considered unsafe .

 Before each use, check that all sensor and alarm ports are clear of any obstructions, i .e . debris or blockage .
 The detector contains a lithium battery that must be disposed of by a qualified recycler . Check local regulations for 

proper disposal .

 DO NOT use IR communications when an explosive atmosphere may be present .
 If you suspect any malfunction or have any technical problems, contact GCT at 1-877-525-0808 .
 The battery may present a fire or chemical burn hazard if mistreated . Do not disassemble, heat above 100°C 

(212°F) or incinerate . Contact Gas Clip Technologies for replacement instructions . Use of another battery may 
present a risk of fire or explosion . 

 Keep new and used batteries away from children .
 Do not expose the detector to sensor poisons such as, but not limited to, alcohol and citrus based products .  

Exposure to poisons may impair the accuracy and/or response time of the detector .

 If suspected sensor poisoning has occurred, recheck the detector (both calibrate and bump test) .
 Caution: Before each day’s usage, sensitivity must be tested on a known concentration of methane equivalent 

to 25-50% of full-scale concentration . Accuracy must be within 0-20% of actual . Accuracy may be corrected by 
calibrating the detector . 

 Any rapid up-scale reading, followed by a declining or erratic reading, may indicate a gas concentration beyond 

upper scale limit which may be hazardous .

 Toute rapide haut de gamme lecture suivie d'une lecture erratique baisse ou peut indiquer une concentration de 

gaz supérieure à la limite superieure de l'échelle qui peut être dangereux .

 Strong Electromagnetic Interference (EMI) may cause incorrect operations .
 Only the combustible gas detection portion of this instrument has been assessed for performance by CSA 

International .

 READ FIRST BEFORE OPERATION

Gas Clip Technologies (GCT) Multi Gas Clip Simple Plus (MGC-S-PLUS) detectors are personal safety devices designed to 
detect the presence of specific toxic gases: carbon monoxide (CO), hydrogen sulfide (H

2

S), oxygen (O

2

) and combustible 

gases/Lower Explosive Limit (LEL) . Before operation, ensure that you have been properly trained on the use of the 
equipment and the appropriate actions to take in the event of an alarm condition . 

Содержание Multi Gas Clip Simple Plus

Страница 1: ...UM MGC S PLUS v1 00a Portable gas detectors you can count on 1 of 18 Multi Gas Clip Simple Plus G A S C L I P T E C H C O M User s Manual...

Страница 2: ...4 Off Mode 4 Alarms 5 Alarm Behavior 5 Default Alarms 5 Detector Menus 6 Main Menu 6 Status Menu 6 Option Menu 7 Adjustable Options 7 Detector Maintenance 9 Battery 9 Bump Testing 9 Manual Bump Test...

Страница 3: ...the detector to sensor poisons such as but not limited to alcohol and citrus based products Exposure to poisons may impair the accuracy and or response time of the detector If suspected sensor poisoni...

Страница 4: ...Ports 4 O2 Sensor 5 Dual Tox Sensor 6 Audible Alarm Port 7 Display 8 Alarm Bar LEDs 9 Maintenance LEDs 10 Infrared IR Communication Window 11 Alligator Clip with Safety Ring 12 Certification Label 13...

Страница 5: ...n Test Mode icon 3 will appear and flash indicating it is time to apply gas The Battery Indicator 4 is displayed as a battery icon which is solid as long as the battery is not low The battery won t ne...

Страница 6: ...r the sensor reading while it is stabilizing and being self tested Once all sensors have completed the warm up and stabilization sequence 65 seconds the detector is fully operational and ready to dete...

Страница 7: ...ry outline only 20 Minutes remaining single beep flash battery outline only 10 Minutes remaining single beep flash battery outline flashes 5 Minutes remaining continuous beep flash every 5 seconds Exp...

Страница 8: ...ill immediately return to normal operation Status Menu The Status Menu is accessed by pressing the Power Menu Button once during the SHOW STATUS prompt The detector will display in the following order...

Страница 9: ...is GAS CLIP TECHNOLOGIES Alarm Limits This option allows you to set sensor low and high alarm limits Each sensor contains separate alarm threshold values that tell the detector when to go into alarm...

Страница 10: ...be adjusted only with the GCT IR Link by Volume CH4 Allows the detector to display gas readings as by volume CH4 rather than LEL The conversion rate can be configured by specifying what the methane co...

Страница 11: ...nual Bump Test Instructions Press the Power Menu Button two separate times in quick succession double tap to access the Main Menu The display will then show in the following order When the screen disp...

Страница 12: ...MGC S Calibration Cap and the detector will return to normal operation after a short period of time The bump due date will be automatically reset based on the new bump date The bump test will be recor...

Страница 13: ...failed a bump test be used During a manual bump test the audible visual vibrating prompting to apply gas will continue for approximately two minutes If gas is not applied within the two minutes then...

Страница 14: ...is detected and as the detector adjusts the calibration parameters Once calibration is complete the detector will display the next calibration date before returning to normal alarming operation If a...

Страница 15: ...first bump test and so on Bump tests are differentiated from normal events when the alarm condition occurs inside of an MGC Simple Clip Dock or when the detector is manually bump tested according to t...

Страница 16: ...ndow lined up approximately 2 3 inches apart 1 Open GCT IR Link Software downloaded for free from www gascliptech com under Resources tab 2 Click on Download Logs icon at the top left corner 3 Select...

Страница 17: ...MGC S CSK Hard sided Carrying Case with Foam Insert 1 ft Sampling Probe Air Stone Particulate Filter MGC S Calibration Cap 10 ft Sampling Hose and 3 ft Calibration Test Hose with Quick Connect Hand A...

Страница 18: ...ox Sensor P N MGC SE 4DT O2 Sensor P N MGC SE O2 Replacement Filters Filter Pack of 10 P N MGC S FILTER 10 Filter Pack of 2 P N MGC S FILTER 2 Various authorized replacement spare parts for Multi Gas...

Страница 19: ...ge Resolution Accuracy T90 H2S 0 100 ppm 0 1 2 ppm 30s CO 0 500 ppm 1 5 ppm 30s Combustible 0 100 LEL 0 1 See Table 1 page 18 30s O2 0 30 vol 0 1 0 7 vol 15s Sensor Type H2S CO O2 Single plug in elect...

Страница 20: ...s and any other applicable recommendations of GCT 2 The buyer promptly notifying GCT of any defect and if required promptly making the product available for correction No goods shall be returned to GC...

Отзывы: