Gas Clip Technologies Multi Gas Clip Simple Plus Скачать руководство пользователя страница 15

Multi Gas Clip Simple Plus User’s Manual

UM MGC-S-PLUS v1.00a 

Portable gas detectors you can count on

13 of 18

DETECTOR RECORDS (LOGS) 

During operation, the detector automatically records all usage activity. These records can be downloaded from the detector 
using the or MGC Simple Clip Dock.

The Event, Bump and Calibration Logs are always downloaded. Partial data logs may be downloaded in order to reduce the 
transfer time. Partial data logs contain approximately 1 week of data. New data log downloads will only contain the data since 
the last download. Full log downloads contain the entire data log, typically at least 2 months’ worth of data. 

Event Log 

The detector stores the last 50 alarm events. These are organized by first in, first out (FIFO) so the 51st event will replace the 
first event and so on. The detector records the specific alarm conditions for each event as follows: 

•  Date and time at the start of the event
•  Duration of the alarm condition
•  Each sensor’s peak alarm status and reading

Bump Log 

The detector stores the last 50 bump tests. These are organized by first in, first out (FIFO) so the 51st bump test will replace 
the first bump test and so on. Bump tests are differentiated from normal events when the alarm condition occurs inside of an 
MGC Simple Clip Dock, or when the detector is manually bump tested according to the Bump Testing section. The detector 
records the bump status for each test as follows:

•  Date and time of bump test
•  If the bump test was performed manually or with the MGC Simple Clip Dock
•  Each sensor’s peak alarm status and reading
•  The result of each sensor’s bump test

Calibration Log 

The detector records the last 50 calibration attempts. These are organized by first in, first out (FIFO) so the 51st calibration will 
replace the first calibration and so on. Each calibration attempt will be recorded as follows:

•  Date and time of calibration
•  If the calibration was performed manually or with the MGC Simple Clip Dock
•  Each sensor’s gas concentration calibrated to
•  Each sensor’s calibration success status

Data Log 

The detector records its current operational status every second. The logging interval cannot be adjusted, but the detector 
compresses the data to reduce the storage and transfer times of redundant records. The typical logging capacity is at least 2 
months of data. The following items are recorded into the log:

•  Date and time
•  Sensor readings and status conditions
•  All user and sensor options
•  Events (i.e. low alarm, high alarm)

Содержание Multi Gas Clip Simple Plus

Страница 1: ...UM MGC S PLUS v1 00a Portable gas detectors you can count on 1 of 18 Multi Gas Clip Simple Plus G A S C L I P T E C H C O M User s Manual...

Страница 2: ...4 Off Mode 4 Alarms 5 Alarm Behavior 5 Default Alarms 5 Detector Menus 6 Main Menu 6 Status Menu 6 Option Menu 7 Adjustable Options 7 Detector Maintenance 9 Battery 9 Bump Testing 9 Manual Bump Test...

Страница 3: ...the detector to sensor poisons such as but not limited to alcohol and citrus based products Exposure to poisons may impair the accuracy and or response time of the detector If suspected sensor poisoni...

Страница 4: ...Ports 4 O2 Sensor 5 Dual Tox Sensor 6 Audible Alarm Port 7 Display 8 Alarm Bar LEDs 9 Maintenance LEDs 10 Infrared IR Communication Window 11 Alligator Clip with Safety Ring 12 Certification Label 13...

Страница 5: ...n Test Mode icon 3 will appear and flash indicating it is time to apply gas The Battery Indicator 4 is displayed as a battery icon which is solid as long as the battery is not low The battery won t ne...

Страница 6: ...r the sensor reading while it is stabilizing and being self tested Once all sensors have completed the warm up and stabilization sequence 65 seconds the detector is fully operational and ready to dete...

Страница 7: ...ry outline only 20 Minutes remaining single beep flash battery outline only 10 Minutes remaining single beep flash battery outline flashes 5 Minutes remaining continuous beep flash every 5 seconds Exp...

Страница 8: ...ill immediately return to normal operation Status Menu The Status Menu is accessed by pressing the Power Menu Button once during the SHOW STATUS prompt The detector will display in the following order...

Страница 9: ...is GAS CLIP TECHNOLOGIES Alarm Limits This option allows you to set sensor low and high alarm limits Each sensor contains separate alarm threshold values that tell the detector when to go into alarm...

Страница 10: ...be adjusted only with the GCT IR Link by Volume CH4 Allows the detector to display gas readings as by volume CH4 rather than LEL The conversion rate can be configured by specifying what the methane co...

Страница 11: ...nual Bump Test Instructions Press the Power Menu Button two separate times in quick succession double tap to access the Main Menu The display will then show in the following order When the screen disp...

Страница 12: ...MGC S Calibration Cap and the detector will return to normal operation after a short period of time The bump due date will be automatically reset based on the new bump date The bump test will be recor...

Страница 13: ...failed a bump test be used During a manual bump test the audible visual vibrating prompting to apply gas will continue for approximately two minutes If gas is not applied within the two minutes then...

Страница 14: ...is detected and as the detector adjusts the calibration parameters Once calibration is complete the detector will display the next calibration date before returning to normal alarming operation If a...

Страница 15: ...first bump test and so on Bump tests are differentiated from normal events when the alarm condition occurs inside of an MGC Simple Clip Dock or when the detector is manually bump tested according to t...

Страница 16: ...ndow lined up approximately 2 3 inches apart 1 Open GCT IR Link Software downloaded for free from www gascliptech com under Resources tab 2 Click on Download Logs icon at the top left corner 3 Select...

Страница 17: ...MGC S CSK Hard sided Carrying Case with Foam Insert 1 ft Sampling Probe Air Stone Particulate Filter MGC S Calibration Cap 10 ft Sampling Hose and 3 ft Calibration Test Hose with Quick Connect Hand A...

Страница 18: ...ox Sensor P N MGC SE 4DT O2 Sensor P N MGC SE O2 Replacement Filters Filter Pack of 10 P N MGC S FILTER 10 Filter Pack of 2 P N MGC S FILTER 2 Various authorized replacement spare parts for Multi Gas...

Страница 19: ...ge Resolution Accuracy T90 H2S 0 100 ppm 0 1 2 ppm 30s CO 0 500 ppm 1 5 ppm 30s Combustible 0 100 LEL 0 1 See Table 1 page 18 30s O2 0 30 vol 0 1 0 7 vol 15s Sensor Type H2S CO O2 Single plug in elect...

Страница 20: ...s and any other applicable recommendations of GCT 2 The buyer promptly notifying GCT of any defect and if required promptly making the product available for correction No goods shall be returned to GC...

Отзывы: