Multi Gas Clip Simple Plus User’s Manual

UM MGC-S-PLUS v1.00a 

Portable gas detectors you can count on

8 of 18

ADJUSTABLE OPTIONS continued

TWA Method

 - The algorithm used to calculate the TWA (Time Weighted Average) can be set to either an average over a moving 

window (OSHA) or as a cumulative average (ACGIH). 

TWA Interval

 - The TWA Interval defines the time frame over which the long-term average is calculated. Default is 8 hours.

STEL Interval

 - The STEL Interval defines the time frame over which the short-term average is calculated. Default is 15 minutes. 

Sensor Enable/Disable

 - Individual sensors can be disabled. A disabled sensor is completely removed from the detector’s display 

for sensor readings, alarm limits and calibrations. This option can be enabled/disabled only with the GCT IR Link.

 Caution:  A disabled sensor will not measure gas or detect alarm conditions .

Bump Interval

 - The Bump Interval controls how often the detector notifies the user to bump test the sensors. The interval can be 

individually adjusted for each sensor.

Calibration Interval

 - The Calibration Interval controls how often the detector notifies the user to calibrate the sensors. The 

interval can be individually adjusted for each sensor. 

Calibration Gas

 - When the detector is calibrated, it scales the sensor readings to match the concentrations of the applied gases. 

The calibration gas concentrations can be adjusted to match the respective levels contained within the gas bottle.
Default is 25ppm H

2

S, 100ppm CO, 18% O

2

 and 50% LEL (2.5% vol CH

4

). This option can be adjusted only with the GCT IR Link.

%-by-Volume CH

4

 - Allows the detector to display gas readings as %-by-volume CH

4

 rather than % LEL. The conversion rate can 

be configured by specifying what the methane concentration is for 50% LEL (typically 2.5% in N. America and 2.2% in EU). 

Language

 - The detector is able to display all its text prompts in any of six languages: English, German, French, Spanish, Italian or 

Portuguese. 

Note: “OL”, “ERROR” and sensor icons (H2S, CO, O2 & LEL) remain the same for all languages

.

Activity LED Period 

-  This option allows the detector to flash a single LED at a specified interval to indicate that the detector

is on. The interval between flashes can be set anywhere from 5 seconds to 120 seconds.

• 

If using GCT IR Link Software - under the Advanced Options section, locate the Activity LED Period box and set the 

desired interval by either clicking on the UP arrow or typing the desired interval directly into the box. To finalize, click 

on the WRITE button at the bottom of the screen. To disable this function, type "0" in the Activity LED Period box and 

press ENTER. The box will display "OFF". Click on the WRITE button to finalize. 

• 

If using GCT Manager Software - place a check on the Activity LED box, then go to the LED Period(s) box and set the 

desired interval by either clicking on the UP arrow or typing the desired interval directly into the box. To finalize, click 

on the WRITE button at the bottom of the screen. To disable this function, type "0" in the LED Period(s) box and press 

ENTER. The box will display "OFF". Click on the WRITE button to finalize.

NOTE: Simply setting the interval in the Activity LED Period box or the LED Period(s) box without clicking on the 

WRITE button will not enable/disable this function on the detector .

Содержание Multi Gas Clip Simple Plus

Страница 1: ...UM MGC S PLUS v1 00a Portable gas detectors you can count on 1 of 18 Multi Gas Clip Simple Plus G A S C L I P T E C H C O M User s Manual...

Страница 2: ...4 Off Mode 4 Alarms 5 Alarm Behavior 5 Default Alarms 5 Detector Menus 6 Main Menu 6 Status Menu 6 Option Menu 7 Adjustable Options 7 Detector Maintenance 9 Battery 9 Bump Testing 9 Manual Bump Test...

Страница 3: ...the detector to sensor poisons such as but not limited to alcohol and citrus based products Exposure to poisons may impair the accuracy and or response time of the detector If suspected sensor poisoni...

Страница 4: ...Ports 4 O2 Sensor 5 Dual Tox Sensor 6 Audible Alarm Port 7 Display 8 Alarm Bar LEDs 9 Maintenance LEDs 10 Infrared IR Communication Window 11 Alligator Clip with Safety Ring 12 Certification Label 13...

Страница 5: ...n Test Mode icon 3 will appear and flash indicating it is time to apply gas The Battery Indicator 4 is displayed as a battery icon which is solid as long as the battery is not low The battery won t ne...

Страница 6: ...r the sensor reading while it is stabilizing and being self tested Once all sensors have completed the warm up and stabilization sequence 65 seconds the detector is fully operational and ready to dete...

Страница 7: ...ry outline only 20 Minutes remaining single beep flash battery outline only 10 Minutes remaining single beep flash battery outline flashes 5 Minutes remaining continuous beep flash every 5 seconds Exp...

Страница 8: ...ill immediately return to normal operation Status Menu The Status Menu is accessed by pressing the Power Menu Button once during the SHOW STATUS prompt The detector will display in the following order...

Страница 9: ...is GAS CLIP TECHNOLOGIES Alarm Limits This option allows you to set sensor low and high alarm limits Each sensor contains separate alarm threshold values that tell the detector when to go into alarm...

Страница 10: ...be adjusted only with the GCT IR Link by Volume CH4 Allows the detector to display gas readings as by volume CH4 rather than LEL The conversion rate can be configured by specifying what the methane co...

Страница 11: ...nual Bump Test Instructions Press the Power Menu Button two separate times in quick succession double tap to access the Main Menu The display will then show in the following order When the screen disp...

Страница 12: ...MGC S Calibration Cap and the detector will return to normal operation after a short period of time The bump due date will be automatically reset based on the new bump date The bump test will be recor...

Страница 13: ...failed a bump test be used During a manual bump test the audible visual vibrating prompting to apply gas will continue for approximately two minutes If gas is not applied within the two minutes then...

Страница 14: ...is detected and as the detector adjusts the calibration parameters Once calibration is complete the detector will display the next calibration date before returning to normal alarming operation If a...

Страница 15: ...first bump test and so on Bump tests are differentiated from normal events when the alarm condition occurs inside of an MGC Simple Clip Dock or when the detector is manually bump tested according to t...

Страница 16: ...ndow lined up approximately 2 3 inches apart 1 Open GCT IR Link Software downloaded for free from www gascliptech com under Resources tab 2 Click on Download Logs icon at the top left corner 3 Select...

Страница 17: ...MGC S CSK Hard sided Carrying Case with Foam Insert 1 ft Sampling Probe Air Stone Particulate Filter MGC S Calibration Cap 10 ft Sampling Hose and 3 ft Calibration Test Hose with Quick Connect Hand A...

Страница 18: ...ox Sensor P N MGC SE 4DT O2 Sensor P N MGC SE O2 Replacement Filters Filter Pack of 10 P N MGC S FILTER 10 Filter Pack of 2 P N MGC S FILTER 2 Various authorized replacement spare parts for Multi Gas...

Страница 19: ...ge Resolution Accuracy T90 H2S 0 100 ppm 0 1 2 ppm 30s CO 0 500 ppm 1 5 ppm 30s Combustible 0 100 LEL 0 1 See Table 1 page 18 30s O2 0 30 vol 0 1 0 7 vol 15s Sensor Type H2S CO O2 Single plug in elect...

Страница 20: ...s and any other applicable recommendations of GCT 2 The buyer promptly notifying GCT of any defect and if required promptly making the product available for correction No goods shall be returned to GC...

Отзывы: