Operation
Operation manual VCpro 600
31
Parameter
Values from/to
Resolution
Default
Unit
Cleaning Time
1…600
1
120
sec
The ventilation valve is opened and, due to the flow rate generated by the pump, the pump
and the pneumatic supply lines are purged in the specified time. The respective device mod-
els are controlled as follows:
VCpro 601 model
vacuum control valve open
VCpro 602 model
maximum motor speed of the vacuum pump connected
ATTENTION
►
The application is ventilated with this procedure.
►
If no other gas type is connected to the inert gas connection of the
ventilation valve, the cleaning is done with air.
6.2.6
System settings
SYSTEM SETTINGS
Language
Display
Control System
Maintenance Timer
Parameter
Values from/to
Resolution
Default
Unit
1
– Language
English
German
French
Spanish
Russian
Chinese
English
NOTE
If you have inadvertently selected the wrong language and you therefore can no
longer navigate, the following process will help:
1.
Reset to factory setting
(= last option under system settings)
2.
Do you really want to reset? = Yes (right) the language will be reset to English.