Operation
Operation manual VCpro 600
29
6.2.3
Operating mode 3 Program
In the operating mode
Program
any pressure profile between 1100…0 mbar can be
programmed and up to 100 cycles run.
PARAMETER
VALUE
Unit
Adjust Pressure Table
1
Line(s)
Number of Cycles
1
#
Vent at Finish
No
Parameter
Values from/to
Resolution
Default
Unit
Pressure Table
1…20
1
1
Line(s)
Number of Cycles
1…100
1
1
#
Vent at Finish
Yes / No
No
A program includes a maximum of 20 setpoints. There is linear interpolation between the
setpoints. The input of the setpoints p and the adjus
tment time ∆t are specified in tabular
form.
∆
t [hh:mm:ss]
p [mbar]
00:00:00
0
Add new line
Parameter
Values from/to
Resolution
Default
Unit
∆t
00:00:00…23:59:59
01:01:01
00:00:00
hh:mm:ss
p
1…1100
1
0
mbar
The set pressure setpoint p is reached in a defined timespan "setting time" ∆t. In the pro-
cess, depending on the previous actual value, the pressure is reduced to the setpoint, held
or increased. There is linear interpolation between the pressure setpoints.