The Reference 600+ is normally provided with an AC line cord suitable for your location. This AC line
cord connects the AC mains to the AC power adapter. If your Reference 600+ has been provided without
an AC line cord, or a cord that is not compatible with your local AC mains socket, obtain a line cord
certified for use in your country. Contact your local Gamry Representative or email to
if you are uncertain what AC line cord to use.
Grounding in the Reference 600+
The circuitry and the metal case of the Reference 600+ are
not
connected to an earth ground. If they
were connected to earth ground, it would compromise the Reference 600+
’s ability to make
measurements in electrochemical cells that contain earth grounded metal. A few examples of such cells
include autoclaves, metallographic stress apparatus, and detectors for capillary electrophoresis.
Most electrochemical cells are isolated from earth ground, so isolation of the Reference 600+ from earth is
not required. In these cases, connection of the Reference 600+ chassis to an earth ground may lower the
noise seen in an electrochemical test. A Chassis Ground binding post on the rear panel of the Reference
600+ makes for easy implementation of this connection. Simply run a wire from this binding post to a
suitable source of earth ground. A black 1.2-meter wire is provided with the Reference 600+ to ease this
connection.
Note this connection of the Reference 600+
to an earth ground is not a “Protective Earth Ground” as
defined in EN 61010. The Reference 600+ is safe in the absence of this connection.
The Reference 600+’s AC Adapter is rated for operation from 100 to 240 V AC, 47 to 63
Hz. It should therefore be useful throughout the world.
Sources of earth ground include
• Most metal water pipes,
• The chassis of
most
electronic apparatus (which are generally earth-grounded), and
• The protective ground terminal of an AC mains power plug.
We recommend that you discuss grounding with an electrical or electronics professional prior to
making this earth-ground connection.
Warning:
Do not connect the chassis ground binding post to any voltage other than
earth ground. An improper connection can create a safety hazard, which could result in personal injury
or death.
Содержание Reference 600+
Страница 2: ...2...
Страница 6: ...4...
Страница 18: ...12 Figure 3 1 Reference 600 Potentiostat Board in Potentiostat Mode Simplified Schematic Block Diagram...
Страница 60: ...54 Figure 7 1 Fast Combination Reference Electrode SCE Platinum White Cell Lead 100 pF to 10 nF Electrolyte...
Страница 70: ...64...
Страница 76: ...Reference 600 Specifications 70...
Страница 80: ...Misc I O Connector 74...
Страница 82: ...Appendix D Auxiliary A D Input Characteristics 76...
Страница 84: ...Appendix E CE Certificate 78...
Страница 85: ...Appendix E CE Certificate 79 Certificate of Conformance...