Reference 600+ Specifications
66
Current to Voltage Converter
Maximum Full Scale Range
600
mA
Note 11
Minimum Full Scale Range
60
600 (after ×100 gain)
pA
fA
Note 11
Voltage across Rm
Typ
120
mV at
full scale
Note 12
Output Voltage (at BNC and ADC in)
Typ
3.0
V
full scale
Input Offset Current
Max
10
pA
Note 13
Range Zero Offset
Max
0.05
% of
range
Note 13
Gain Tolerance
600 mA to 6 nA ranges
600 pA range
60 pA
Max
0.2
0.75
1.5
% of
reading
Note 13
Zero drift
Typ
0.03
% FS/°C
Note 14
Bandwidth
600 mA to 600
A ranges
60
A range
6
A range
Typ
> 10
> 1.5
> 0.15
MHz
Note 15
Current Measurement
Resolution
Typ
0.00333…
% FS
/bit
Offset Range
Typ
100
% of
range
Post- Offset Gain
Typ
1×, 10×, 100×
Accuracy
Typ
Dominated by current-to-voltage error
(see above)
Potentiostatic Mode
Applied Voltage Range
Min
11
V
Notes 5, 6
Accuracy
DC zero offset
Gain
Max
1
0.2
mV
% setting
Note 16
DC Bias
Typ
8
V
Scan DAC ranges
Typ
6.4,
1.6,
0.4
V
Drift
Max
< 20
V/°C
Note 17
Noise and Ripple
1 Hz to 1 kHz
1 Hz to 200 kHz
Typ
10
20
V rms
Note 18
Содержание Reference 600+
Страница 2: ...2...
Страница 6: ...4...
Страница 18: ...12 Figure 3 1 Reference 600 Potentiostat Board in Potentiostat Mode Simplified Schematic Block Diagram...
Страница 60: ...54 Figure 7 1 Fast Combination Reference Electrode SCE Platinum White Cell Lead 100 pF to 10 nF Electrolyte...
Страница 70: ...64...
Страница 76: ...Reference 600 Specifications 70...
Страница 80: ...Misc I O Connector 74...
Страница 82: ...Appendix D Auxiliary A D Input Characteristics 76...
Страница 84: ...Appendix E CE Certificate 78...
Страница 85: ...Appendix E CE Certificate 79 Certificate of Conformance...