65
Appendix A:
Reference 600+ Specifications
All specifications are at an ambient temperature of 25°C, with the Reference 600+ powered using the
external power supply shipped with the unit, a standard shielded 60 cm cell cable, and the cell enclosed in
a Faraday shield. All specifications are after software calibration.
A numbered note qualifies many of the specifications. Many of these notes describe the method used to
measure a specification. The notes are found at the end of this appendix.
All specifications are subject to change without notice.
Control Amplifier
Compliance Voltage
Min
22
V
Note 1
Output Current
Min
600
mA
Note 2
Unity Gain Bandwidth
Typ
980, 260, 40, 4.0, 0.4
kHz
Notes 3, 4
Slew Rate
Typ
10, 4.5, 0.6, 0.06, 0.006
V/
s
Notes 3, 4
Differential Electrometer
Max Input Voltage
Min
9
V
Note 5
Input Current
Max
10
pA
Notes 5, 6
Input Resistance
Differential (between inputs)
Common Mode (input to ground)
Typ
100
1
T
Ω
Note 7
Input Capacitance
Differential (between inputs)
Common Mode (input to ground)
Typ
0.2
12
pf
Note 7
Bandwidth (
–
3 dB)
Min
15
MHz
Note 7
CMMR
DC to 100 kHz
100 kHz to 1 MHz
Min
80
60
Note 8
Voltage Measurement
A/D Full Scale Ranges
Typ
12.0,
3.0
V
Note 9
Resolution
Typ
400, 100
V/bit
Zero Offset Error
Max
1
mV
Note 10
Gain Error
Max
0.2
%
Note 10
Offset Range
Typ
10
V
Post- Offset Gain
Typ
1×, 10×, 100×
Содержание Reference 600+
Страница 2: ...2...
Страница 6: ...4...
Страница 18: ...12 Figure 3 1 Reference 600 Potentiostat Board in Potentiostat Mode Simplified Schematic Block Diagram...
Страница 60: ...54 Figure 7 1 Fast Combination Reference Electrode SCE Platinum White Cell Lead 100 pF to 10 nF Electrolyte...
Страница 70: ...64...
Страница 76: ...Reference 600 Specifications 70...
Страница 80: ...Misc I O Connector 74...
Страница 82: ...Appendix D Auxiliary A D Input Characteristics 76...
Страница 84: ...Appendix E CE Certificate 78...
Страница 85: ...Appendix E CE Certificate 79 Certificate of Conformance...