CADIAX compact
56
Press the foot switch to start recording the selected
movement and have the patient carry it out within the fol-
lowing 4.5 seconds.
The start and the end of the recording are signaled
acoustically. During the recording, the progress will be
displayed and the curve will be drawn in real-time.
Once the recording is finished, it will be displayed in the
recording window and you can proceed with the next
one.
Check a curve for its quality immediately after the recording. To do so, check the starting point
of the movement track relative to the reference position (i.e. the origin of the coordinate system),
as well as smoothness of the curve line. Ideally, the movement starts exactly in the zero point
and doesn't show any distortion artifacts of any kind.
Registrations with obvious distortions should not be accepted. In such cases, clean the flags
and styli thoroughly and repeat the recording. If this doesn't prevent artifacts from appearing,
cancel the registration and contact an authorized GAMMA service department.
In addition, movement tracks that differ from the expected shape should be treated skeptically
and have to be verified with further registrations.
Please note that the designations mediotrusion right and mediotrusion left always refer to the
condyle of the patient that is carrying out the mediotrusive movement. Mediotrusion right there-
fore means a mandibular (i.e., of the chin) movement to the left side and vice versa.