Recording TMJ movement
41
3.5
Mounting the kinematic facebow
3.5.1
Condylograph
3.5.1.1
Preparing the upper face bow
Before mounting the Condylograph on the patient, the
forehead bracings of the upper face bow need to be
pulled back.
Slightly loosen the side arm clamp screws, so that the
side arms are movable in sagittal (upper clamp screw)
and transversal (lower clamp screw) direction.
Loosen and remove the orbital pointer.
The upper face bow is now ready to be mounted.
3.5.1.2
Mounting the upper face bow
Adjust the width of the face bow in a way, that the side
arms are in contact with the skull above the ears evenly
without tension. The face bow should be adjusted sym-
metrically.
Fixate the flag adapters of the extension set with the re-
spective screws on the side arms.