Hex
command
Description
Data field Response
Data/Response description
FF
MASTER RESET
(legacy)
Executes a
complete software
reset (legacy MPC
support command).
No response due to
initialization of controller
91
SET ARC DETECT
Enables/disables
arc detection.
"YES" or
"NO"
92
GET ARC DETECT
Reads the current
arc detection
setting.
"YES" or
"NO"
0A
READ CURRENT
Reads pump
current.
None or
"1"
X.XE-XX
AMPS
0.1E-09 = HV OFF STATUS
0B
READ PRESSURE
Reads pump
pressure.
None or
"1"
X.XE-XX UUU UUU is pressure units (Torr,
MBR, or PA) 0.1E-10 = HV
OFF STATUS
0C
READ VOLTAGE
Reads pump
voltage.
None or
"1"
XXXX
XXXX indicates the voltage In
volts
0D
GET SUPPLY
STATUS
Reads various
conditional
messages displayed
on the front panel.
None or
"1"
Various
See Display Messages
section of this manual
0E
SET PRESS UNITS
Specifies the default
pressure units.
U
U is pressure units
U = T (Torr), M (Mbar), P
(Pascal)
11
GET PUMP SIZE
Reads pump size in
liters per second.
None or
"1"
ssss L/S
ssss is pump size
12
SET PUMP SIZE
Sets pump size in
liters per second.
ssss
ssss is pump size
1D
GET CAL FACTOR
Reads the
calibration factor
that modifies
pressure.
n.nn
n.nn range is 0.00 – 9.99
1E
SET CAL FACTOR
n.nn
n.nn range is 0.00 – 9.99
04/2021 - ©Gamma
Page 44
900026_F