POWER-LOSS RESTART
You can choose yes or no to have both pumps start automatically after a power loss.
1.
Press the
Menu
key to cycle through the menu until you reach CONFIG.
2.
Press
ENTER
to step to the submenu and press the
Menu
key to cycle to POWER-
LOSS.
3.
Press
ENTER
.
4.
An arrow points to the bottom
Display Select
key to show it is active. Press it to
toggle between Yes and No.
1
2
3
4
5
6
7
8
Set Point
Set Point
Supply 1
Supply 2
Display
Select
Display
Select
arrow shows
bottom key is active
Options:
- NO (manual restart)
- YES (automatic restart
press to toggle
between options
5.
When you reach the desired option, press ENTER to save your selection.
SERIAL ADR
You can choose which serial address to use for serial communications. Default is 5.
1.
Press the
Menu
key to cycle through the menu until you reach CONFIG.
2.
Press
ENTER
to step to the submenu and press the
Menu
key to cycle to SERIAL
ADR.
3.
Press
ENTER
. The bottom display shows the current address followed by a flashing
sign to indicate that you may key in a new address (0 – 255).
1
2
3
4
5
6
7
8
Set Point
Set Point
Supply 1
Supply 2
Display
Select
Display
Select
flashing sign
Options:
0 through 255
4.
First use the backspace (
←
) key to overwrite, and then enter the desired address with
the numerical keys.
5.
Press
ENTER
to save the changes.
BAUD RATE
You can select the baud rate to use: 9600, 4800, 2400, 1200, or 300. The default is 9600. Baud
rate can be decreased to allow communication through longer serial cables.
1.
Press the
Menu
key to cycle through the menu until you reach CONFIG.
2.
Press
ENTER
to step to the submenu and press the
Menu
key to cycle to BAUD
RATE.
3.
Press
ENTER
again to display the rate.
4.
An arrow points to the bottom
Display Select
key to show it is active. Press it to
cycle through the baud rate options — 9600, 4800, 2400, 1200, or 300.
Digitel MPC Manual
Page 34 of 72