0
ER
Operate
TSP
Menu
1.
Press
Menu
to step through each top menu selection.
2.
Use the
ENTER
key to step to a sub menu.
3.
Then use
Menu
to step through each submenu selection.
4.
Press
ESC
to step back up to the top menu.
Table III-3. Menu Hierarchical System.
Menu Structure
Description
PUMP1 SIZE
PUMP2 SIZE
Specifies pump size in liters per second that is connected to each supply. This value must
be either 0 (no pump connected) or between 10 and 1200. Default is 0.
PREFRENCES:
The following sub-menus are available in the PREFRENCES menu.
PRES UNITS
Selects pressure units to use: Torr, mBar, or Pascal. Default is Torr.
PUMP DSPLY
Selects voltage, current or pressure for each display. Default is voltage.
ANALOG 1 OUT
MODE
Selects analog current/pressure output mode for supply 1.
ANALOG 2 OUT
MODE
Selects analog current/pressure output mode for supply 2.
CONFIG:
The following sub-menus are available in the CONFIG menu.
FAN ON/OFF
You can set the fan to off or full speed.
POWER LOSS
If yes is chosen, both pumps start up automatically upon power up.
SERIAL ADR
Specifies which serial address to use for the Physical Electronics Serial Communications
Protocol. Default is 5.
BAUD RATE
Baud rate of 9600 is available.
SET CLOCK
Allows you to set the time and date of the real time clock.
SPLY1 SIZE
SPLY2 SIZE
Indicates what supply option is installed for supply 1 and supply 2: medium, large, or none.
PUMP 1 CAL
PUMP 2 CAL
Pumps have a calibration factor that is multiplied by pressure to increase or decrease it.
Acceptable factors are between 0.00 and 9.99. Default, 1.00.
SUPPLY1 KV
SUPPLY2 KV
Indicates what output voltage the high voltage transformer is set to (for supply 1 and 2), 5600
or 7000.
PRM FACTOR
Informs the firmware of line voltage setting, 120 or 240.
DIAGNOSTCS:
Most of the following diagnostics can be used in the field.
REVISION
Displays the current firmware revision.
ERROR
CODE Display 1 shows last error code for pump 1 and display 2, pump 2.
DISPLY/KEY
Tests all displays, LEDs, and keys.
CALIBRATION A self-calibration of the current offset for both supplies.
RST
DFLTS
Resets most options to their default values.
LINE
FREQ
Indicates the present line frequency.
Setting Pump Size
In order to start the pump, you need to specify the pump size (between 10 and 1200 in liters per
second). Use the following procedure
Digitel MPC Manual