WRO 300 H
32
Regular maintenance
HCEN12481-02/2012
11
Keep the wand in the container and disconnect the wand by
turning the connector a quarter of a turn and gently pull it
out. Let the disinfectant liquid from the line and wand flow
back into the container. Then clamp the line. Also check that
correct amount of disinfectant has been consumed.
12
Remove the disinfection wand from the container and flush
with water.
13
The Chemical disinfection procedure will now continue
automatically with
DWELL PERIOD
followed by
RINSE
.
The remaining time to complete the Chemical disinfection
procedure is also displayed.
14
When the Chemical disinfection program is completed the
display will show
PERFORM RESIDUAL TEST
. The indi-
cation
will remain steady lit and the
will flash.
NOTE
The conductivity value is checked during the final five minutes of the
rinse phase. If the value is above the Conductivity notification limit
(refer to
“Conductivity monitoring” on page 16
) when the preset rinse
time is finished it is prolonged 30 minutes. If the value decreases
below the Conductivity notification limit the prolonged rinse finishes.
If the level still is above the Conductivity notification limit when the
prolonged rinse is finished the alarm
403 HIGH CONDUCTIVITY
RINSE
appears.