WRO 300 H
26
Regular maintenance
HCEN12481-02/2012
3.1.1
Heat disinfection
Heat disinfection of the WRO 300 H, including product water
loop.
Procedure
1
Press
until
SELECT
is shown in the display and
starts to flash.
2
From
SELECT
, press
to enter the list of available proce-
dures.
3
Select
HEAT
by using the
.
4
Initiate the Heat disinfection by pressing
until the light
goes on.
Stop of Heat disinfection procedure
•
Press
until light goes on, to stop Heat disinfection proce-
dure. An Active Cooldown is initiated. Then the
WRO 300 H goes to Standby.
More information of the settings of Active Cooldown refer
to
Service Manual
.
NOTE
The dialysis machine shall be turned off when the Heat disinfection
procedure is performed.
NOTE
The Heat disinfection procedure can also be initiated by the time
channels. Further information of time channel settings refer to
Service Manual
.
NOTE
If power failure occurs during Heat disinfection procedure
301
INSUFFICIENT HEAT
is displayed when the WRO 300 H restarts.
Press
until the WRO 300 H either goes to Standby or an Active
Cooldown is initiated if it is preset.