Technical data
HCEN12481-02/2012
Water supply
Input
Min. 3.0 l/min required
Feed Water Pressure
During operation:
0.15 to 0.3 MPa (20 to 45 PSI)
(There is an inlet flow of water into
the WRO unit)
Maximum design pressure:
0.8 MPa (115 PSI)
If the feed water pressure exceeds
0.3 MPa (45 PSI), a flow regulator,
order no W522001001, should be
installed.
Feed Water Temperature
+ 5 to + 30 °C
Feed Water Quality
Potable water should be used.
Particle filtration shall be used
(< 5
μ
m).
Carbon filtration shall be used if
water is chlorinated, see limits
below.
Membrane life expectancy might
be reduced if the monitor is ope-
rated outside the following limits.
Hardness
< 0.3 °dH (6 ppm as CaCO
3
)
Iron
< 0.1 mg/l
Manganese
< 0.1 mg/l
Turbidity
< 1 JTU
Total dissolved salts
(TDS)
< 1500 mg/l
Fouling index (silt density
index)
< 5
Chlorine (total)
< 0.1 mg/l
Values