2.
Remove Allen screws and / or air purge connectors.
Pic
tu
re
2
Pic
tu
re
3
3.
Screw off lamp arm.
4.
Shift the lamp arm until you have access to the
electrical/cable connections.
Pic
tu
re
4
5.
Loosen all wires of the lamp cable by using the 2 mm
flat blade screwdriver.
Remove lamp arm and cable carefully.
6.
Loosen Allen screws (3x) by using the 1.5 mm
hexagonal spanner.
7.
Remove lamp assembly carefully.
Important
: See Pictures 3, 4 and 5. We recommend removing
the lamp assembly and performing the lamp
replacement procedure in a dry and clean location.
Pic
tu
re
5
Pic
tu
re
6
8.
Remove screws by using the medium size Phillips
screwdriver.
9.
Carefully remove the printed circuit board including
cable and socket from the measurement lamp.
Pic
tu
re
7