Galvanic Applied Sciences USA, Inc.
101 Billerica Ave, Bldg. 5, Suite 104.
North Billerica, MA. 01862
Tel: 978-848-2701
Fax: 978-848-2713
Email: [email protected]
13
Pic
tu
re
8
10.
Remove screw joint carefully by using the special
Monitek lamp spanner wrench..
11.
Pull out the measurement lamp carefully by using the
small needle nose pliers.
Important
: Avoid extreme force in pulling out the lamp.
Pic
tu
re
9
Pic
tu
re
10
12.
Adjust the pins of the new lamp before installation of
the new measurement lamp by using the lamp socket.
This will allow an easy final assembly of the
components
13.
Remove plug from the measurement lamp and push the
lamp into the barrel.
Pic
tu
re
11
Pic
tu
re
12
14.
Screw in the screw joint and tighten it by using special
Monitek lamp spanner wrench.
15.
Push the plug carefully onto the contact pins of the
lamp.
Pic
tu
re
13