Before installation and start-up
Installation guidelines
•
The sensor is manufactured according to the customer’s application (variable line size, flange
type, cleaning jets, gasket material etc.).
•
It is recommended to run the calibration of the system before installation of the sensor.
•
The location / installation of the sensor should be in a vertical standpipe.
•
The process pressure should never exceed the specification of the delivered sensor.
•
The process temperature should never exceed the specification of the delivered sensor.
•
Avoid air and gas bubbles inside the sensor, they cause disturbances. Air and gas bubbles will
cause noise and drift of the measurement signal. (The air bubbles are not expected at pressures
upwards of 2 bar in aqueous solutions).
•
In case the process temperature should fall under the dew point or rise above 85 °C purge the
sensor optic housings with dry instrument air (approx. 10 l/h). Condensed water and excessive
temperatures can damage the sensor and cause inaccurate measurements.
•
Due to potential noise problems it is recommended not to extend the sensor cables.
•
Due to potential noise problems use original Monitek sensor cables only.
•
In case the optional cleaning jets are being used, make sure the pressure of the cleaning fluid is at
least 50% higher than the process pressure.
Danger:
Exceeding the specified maximum pressure and /or the specified maximum temperature
will cause a very high safety risk.
Please read the additional safety instructions before installation and start-up.
Page 5 and page 6 !