16.
Put the printed circuit board to the barrel and screw on
by using the medium sized Phillips screwdriver.
17.
Reconnect the wires of the lamp cable and perform a
lamp test.
Pic
tu
re
15
Pic
tu
re
16
Important
: Please pay attention to the position of the optical
assembly. The three grooves on the outside of the
barrel must be in line with the three Allen screws.
18.
Push the optical assembly into the holder.
19.
Tighten Allen screws (3x)
20.
Push lamp arm to the thread.
21.
Tighten lamp arm.
Pic
tu
re
17
Pic
tu
re
18
22.
Tighten cable gland
23.
Screw in Allen screws and / or air purge connectors
24.
Functional test and eventual calibration.
Pic
tu
re
19