![Galvanic Applied Sciences 962-AGX Скачать руководство пользователя страница 91](http://html1.mh-extra.com/html/galvanic-applied-sciences/962-agx/962-agx_operation-manual_3294067091.webp)
Revision 1
91
9/10/2019
Section 10
Specifications
Performance Specifications
Auto Calibration
User Selectable Frequency
Zero Drift
< 0.25% of full scale per day (based on
autozero frequency of once/hour
Sensitivity
+/- 2.0 % of Full Scale
Repeatability
+/- 2 % of Full Scale
Response Time
Analyzer – Near Instantaneous
Total System – Sample Line Dependent
Physical Specifications
Size
Mounted on Frame
Outside Dimensions 57” L x 42”H x16” D
(145 cm x 107 cm x 41 cm)
Weight
250 lbs (114 kg), depending on options
Control Cabinet
30” H x 24” W x 12”D
76 cm x 61 cm x 30 cm
Services Required
Electrical
110/220 V, 1 phase, 50/60 Hz, 15 A
Instrument Air
< 15 SCFM at 80 psi
Nitrogen
Bottled Regulated
Area Classification
Class 1, Division 2, Groups C and D.(Type Z Pressurized)
All specifications are subject to change as part of our ongoing product
improvement program.