Revision 4
71
August 02, 2019
c) Loosen the fluid holder locking screw (s) on the left-side of the fluid holder using a 7/64"
ball driver.
d) Lift the fluid holder from the base by grasping the fluid holder filler plug. Note the
approximate level of
Anti-Solarant Solution
remaining in the chamber (it should be about
75
%).
e) Remove the filler plug from the top of the fluid holder.
f) Dump the old
Anti-Solarant Solution
out.
The
Anti-Solarant Solution
is not hazardous to the environment.
g) Rinse the chamber with a small amount of new
Anti-Solarant Solution
. Shake and empty.
h) Fill the fluid holder with new
Anti-Solarant Solution
. Tilt and tap the fluid holder to remove
any air bubbles that may form.
Take care not to touch the
Anti-Solarant Solution
directly as the oils from
your hands will contaminate the solution.
i) Replace the fluid holder filler plug.
j) Replace the fluid holder into the bracket. Orient the fluid holder so that the face with the
word
LAMP
engraved on it is towards the UV Lamp Supply.
DO NOT touch the fluid holder windows as dirt, grease and human skin oil
will affect their performance.
k) Tighten the fluid holder locking screw (s) on the left-side of the fluid holder using a 7/64"
ball driver.
l) Reinstall the light shield and tighten the light shield retaining screws on the bottom right-
side of the fluid holder.
m) Remove the protective plastic cap from the fiber and reconnect the
Source
optical fibre.
n) Reoptimize the spectrometer energy level as described in Section 6.4
Содержание 943-TGX
Страница 48: ...Revision 4 48 August 02 2019 Figure 4 13 Network Sub Panel...
Страница 94: ...Revision 4 94 August 02 2019 Figure 9 3 Control Cabinet Detail...
Страница 96: ...Revision 4 96 August 02 2019 Figure 9 5 AC Wiring Diagram...
Страница 97: ...Revision 4 97 August 02 2019 Figure 9 6 DC Signal and Wiring Diagram...
Страница 98: ...Revision 4 98 August 02 2019 This Page Intentionally Blank...