Revision 4
43
August 02, 2019
h)
Cell Length
- Indicates the cell length of the measuring cell. This value is used in the
calculation of the gas concentrations.
i)
Fixed Temperature
- To select
Fixed Temperature,
click on the e
Fixed Temperature
on/off
field using the
Field Next
button on the keypad. Use the
Enter
key to toggle the x
on and off in this field. If the x is present, the analyzer will use the temperature indicated
here when calculating the concentration of the various gases If the x is not present, the
analyzer will use the live value for cell temperature when calculating the concentration of
the various gases.
j)
Fixed Pressure
- To select
Fixed Pressure
navigate to the
Fixed Pressure on/off
field
using the
Field Next
button on the keypad. Use the
Enter
key to toggle the x on and off in
this field. If the x is present, the analyzer will use the pressure indicated here when
calculating the concentration of the various gases. If the x is not present, the analyzer
will use the live value for cell pressure when calculating the concentration of the various
gases.