VACSOUND ECO – Operating and maintenance instructions
27
1. Place a tank for the waste oil under the oil drain plug (79).
NOTE
: You are advised to empty the pump before it cools down completely so that the
oil flows out more easily.
NOTE
: The pump contains up to 1 litre of oil - make sure the tank you use for the waste
oil can hold this amount.
2. Open the oil drain plug slowly; let the oil flow out by opening the fill plug (78) located at
the top of the pump. In the meantime, clean the inside of the drain plug (magnetic
plug).
3. After closing the drainage hole with its plug, make the pump perform a number of turns,
giving it a brief current pulse, so as to let residual oil in the pump chambers flow out.
Then drain off the residual oil again.
4. After all the oil has been drained, close the drain plug tight; take the tank of waste oil
away and clean thoroughly, using rags of cotton or other suitable material to dry.
5. Clean thoroughly the pourer lip of the new oil tin; pour the oil slowly into the pump
through the filler plug.
WARNING
: Monitor through the sight glass at the front of the pump that you are not
exceeding the maximum oil level permitted.
CAUTION
: For disposal of waste oil, it is strictly necessary to respect the currently
applicable legislation on this matter.
6. To achieve a good vacuum level again, run the pump for one hour with the gas ballast
open.