TEMPVIEWER
3600
7
0%
30%
60%
80% 100%
MANUAL
Mixing images
The Tempviewer3600 can combine visual images in 5 ways with thermal imagery.
The heat image can be displayed 0, 30, 60, 80 or 100%, respectively, over the visual
image. To choose how much% of the heat image is visible on the visual image, use
the Navigation keys
|
and
}
.
MENU
∙ To adjust settings in the menu, briefly press the “Power / Menu key”. The menu will
be visible. Use the Navigation buttons
~
and
to scroll through the menu with the
yellow selection box.
∙ To adjust a menu function, bring the yellow selection box to the desired item and
press the “Select / Enter” key. Choose the desired setting with the Navigation but-
tons
~
and
And confirm with the “Select / Enter key”.
∙ Press the “Power / Menu key” to exit the menu.
a. - Unit (°C / °F)
The Tempviewer3600 can measure in both °C and °F. To switch between these units,
press the “Power / Menu key”. Move with the Navigation buttons
~
and
to the “Tem-
perature Unit (°C / °F)” option
(a)
and confirm with the “Select / Enter” key. Choose
with the Navigation buttons ~ and ?? The desired unit and confirm with the “Select /
Enter key”
b. - Background temperature of reflective substrate (Low emissivity)
The temperature of the background can be set from 0 °C to 36 °C. This is especially
necessary as temperature compensation for shiny surfaces (low emissivity). Heat or
cold peaks of and around the object can affect the surface temperature and measu-
rement accuracy of the measured object. This becomes especially clear when the
radiation coefficient of the measured object is low. In these situations, only set the
temperature of the reflective background to get the measured result as accurate as
possible. In all other cases, use a temperature of 0 °C
(recommended)
.