To work
∙ Insert 4x AA batteries into the device. Consider the polarity. This is indicated in the
device.
∙ Press 5 sec. On the “Power / Menu key” to turn on the unit.
∙ During the measurements there is a central mark to see. The measured tempera-
ture of this point is displayed in the upper left corner of the screen. The emissivity
value is shown in the upper right corner. At the bottom left, the minimum and maxi-
mum values are displayed within the measurement range. If desired, this minimum
and maximum value can be displayed with a marker on the displayed thermal /
visual image. (See min. / Max. Value)
∙ To save images on the MicroSD card, aim the Tempviewer3600 on the surface to
measure and press the “Recording key.” “Store photos yes no” will appear on the
screen. Press the “Power / Menu key” to save the image, press the “Select / Enter
key” if you don’t want to save the image. If “FULL” appears at the bottom of the
screen, you must clear the MicroSD card.
∙ To switch off the device, press the “Power / Menu key” for 5 seconds until the unit
starts counting down.
Removing noise
Does the image display noise? Then start the device and make sure the sensor is clo-
se to the work surface. Press 5 seconds on the Navigation key
~
until “Adjust” appears
on the screen. This process takes about 30sec.
Led light
Tempviewer3600 features an LED light. This can be activated by pressing the Record
key (10) for about 5 seconds. Repeat this step to turn off the LED light again.