background image

TEMPVIEWER

3600

 

5

MANUAL

How big is the measuring area?

The further you are of the object whose temperature you want to measure, the gre-

ater the measurement area. The larger the measuring surface, the less the accuracy. 

We therefore recommend using the camera as close as possible to the objects to be 

measured.

The applicable distance is 50cm. 

This thermal camera has a measurement ratio of 10: 6. This means that at a measu-

rement at a distance of 100cm of the object, the measuring surface is approximately 

60x60cm in size. 

GOOD TO KNOW…

 ∙ An infrared thermometer must adapt to ambient temperature. This may take up to 

30 minutes to adjust to high temperature fluctuations. So wait a few minutes bet-

ween measuring a hot and cold object.

 ∙ An infrared thermometer can not be measured through transparent surfaces (eg. 

glass). He will then measure the temperature of the glass.

 ∙ Steam, dust, smoke ... make measurements with an infrared thermometer less accu-

rate to unreliable..

Содержание TEMPVIEWER3600

Страница 1: ...TEMPVIEWER3600 MANUAL...

Страница 2: ...that the bag and instrument are dry otherwise condensation may occur in the device Make sure that the windows are free of dirt and clean them using a soft cloth and a glass cleaning product only Alway...

Страница 3: ...tton 3 Battery compartment 4 Selection Enterbutton 5 Micro SD card 6 Navigation 7 Infrared sensor 8 Camera 9 LED light 10 Record button 11 Thread 1 4 Menu a Temperature unit C F b Background temperatu...

Страница 4: ...average emission power to a black radiator at the same temperature Emission is for most materials between 0 10 and 1 00 Materials with low emissivity 0 60 emit little energy typi cally for materials...

Страница 5: ...t ratio of 10 6 This means that at a measu rement at a distance of 100cm of the object the measuring surface is approximately 60x60cm in size GOOD TO KNOW An infrared thermometer must adapt to ambient...

Страница 6: ...roSD card aim the Tempviewer3600 on the surface to measure and press the Recording key Store photos yes no will appear on the screen Press the Power Menu key to save the image press the Select Enter k...

Страница 7: ...u a Unit C F The Tempviewer3600 can measure in both C and F To switch between these units press the Power Menu key Move with the Navigation buttons and to the Tem perature Unit C F option a and confir...

Страница 8: ...1 00 Look for the correct value in the table In case of doubt set the emissivity value to 0 95 To set the emissivity press the Power Menu key Move with the Navigation buttons and To the Emissivity Val...

Страница 9: ...to 25 000 images can be saved The menu shows the number of saved pictures To view saved images press the Power Menu key and move with the Navigation keys and To Saved pictures i and confirm with Selec...

Страница 10: ...w to counteract them The product is not to be operated until the user has been instructed on how to work with it LIMITS OF USE Environment Suitable for use in an atmosphere appro priate for permanent...

Страница 11: ...cted to mechanical shock for example blows or falling the product may be damaged or peo ple may sustain injury When setting up the product make sure that the accessories are correctly adapted fitted s...

Страница 12: ...al months to prevent damage If your instrument does not reach the re quired tolerance it should be returned to your service centre or to your reseller for service Repairs carried out by unau thorised...

Отзывы: