5 - 7
ECG
Phenomenon
Cause
Solution
“Electrode?” message
is displayed.
Lead cable is off.
Check the connection between the
lead cable and the LX-8100/
LX-8100(G).
Check the connection between the
lead cable and the electrode.
Faulty lead cable.
Replace the ECG cable with a new
one.
Electrode is peeling off.
Replace the electrode with a new
one.
Polarization potential of the
electrode is too high.
Replace the electrode with a new
one.
ECG waveform
contains noise.
Electrode gel is dry.
Replace the electrode with a new
one.
Electrode is peeling off.
Electric blanket is used.
Cover the electric blanket with a
shield cover.
AC filter setting of the receiving
monitor is OFF.
Set the AC filter to ON.
Respiration waveform
cannot be measured.
Electrode gel is dry.
Replace the electrode with a new
one.
Electrode is peeling off.
The positions of the electrodes
are improper.
Attach the electrodes where the
respiration waveform can be
measured appropriately.