3 - 2
Procedure for Optional Settings
The following settings are available for the LX-8100/LX-8100(G) depending on the usage and condition
of the patient.
These setting changes can be performed on the Preset screen.
Items
Selection
Default
Backup
Channel
LX-8100 Transmitter
Channel (CH)
<For Europe>
One from following
channels.
3305 to 3379
3400 to 3462
3400
○
LX-8100 Transmitter
Channel (CH)
<For USA>
One from following
channels.
0801 to 0879
0900 to 0979
1000 to 1079
1100 to 1179
1200 to 1279
1300 to 1379
1100
○
LX-8100(G) Transmitter
Channel (CH)
<For USA>
One from the following
channels.
9501 to 9539
9600 to 9639
9700 to 9739
9800 to 9839
9900 to 9938
2701 to 2739
2800 to 2839
2900 to 2918
2921 to 2939
3000 to 3039
3100 to 3118
9501
○
Group ID
One from 00 to 63
00
○
ECG
Time Constant
(Time Const.)
0.4 sec, 0.1 sec
0.4 sec
○
Detection Sensitivity of
Pacemaker Pulse
(Pace Sens.)
Low, Mid, High
Mid
○
QRS Detection
(QRS Width)
Wide, Narrow
Wide
○
RESP
Respiration Detection Signal
(Display)
ON, OFF
ON
○
SOUND
SOUND
ON, OFF
ON
○
Display
Display Brightness
(Brightness)
8 levels
5
○
Turn Off Display Time
(Display OFF)
1 min, 3 min, 10 min, OFF
3 min
○
Battery (Europe only)
Battery
AA, Ni-MH