Compliance to the Electromagnetic Immunity
The LX-8100 is intended for use in the electromagnetic environment specified below. The customer or
the user of the LX-8100 should assure that it is used in such an environment.
Basic EMC standard or test method
Immunity test levels
Electrostatic discharge
IEC 61000-4-2
±
8kV contact
±
2kV,
±
4kV,
±
8kV,
±
15kV air
Radiated RF EM fields
IEC 61000-4-3
10V/m
80MHz to 2.7GHz
1kHz
80%AM
Proximity fields from RF wireless
communications equipment
IEC 61000-4-3
Refer to the following table.
Conducted disturbances induced by RF
fields
IEC 61000-4-6
3V
0.15MHz to 80MHz
1kHz
80%AM
6V
0.15MHz to 80MHz
(in ISM bands between 0.15 MHz and 80MHz)
1kHz 80%AM
Rated power frequency magnetic fields
IEC 61000-4-8
30A/m
50Hz
Immunity test specifications for RF wireless communications equipment.
Test frequency
(MHz)
Modulation
Maximum power
(W)
Distance
(m)
Immunity test level
(V/m)
810, 870, 930
PM, 18Hz
2
0.3
28
1720, 1845, 1970
PM, 217Hz
2
0.3
28
2450
PM, 217Hz
2
0.3
28
5240, 5500, 5785
PM, 217Hz
0.2
0.3
9